A simple equation to correct for gadolinium interference on plasma selenium measurement using inductively coupled plasma mass spectrometry
- 3 May 2020
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 57 (3) , 234-241
Abstract
Background: The measurement of selenium in human plasma is useful to assess deficiency or toxicity. The presence of gadolinium in clinical samples following administration of certain contrast agents used for magnetic resonance imaging can cause a significant positive bias in selenium results when measured using quadrupole inductively coupled plasma mass spectrometry (Q-ICP-MS). Methods: A mathematical equation to correct for gadolinium interference was assessed using both patient samples and commercial quality control/external quality assurance (QC/EQA) materials spiked with gadolinium. Samples were analysed using an Agilent 7900 ICP-MS operated in ‘narrow peak’ (half-mass) mode. Accuracy was evaluated by comparing corrected selenium results with target concentrations. Results: Corrected results were found to be accurate at all gadolinium concentrations tested (2, 4, 10 and 20 mg/L). Average recoveries ranged from 97.4 to 106.5%. Results for QC/EQA materials were within specified target ranges. Within-run imprecision was <3%, and between-run imprecision was <4.3%, demonstrating robustness. Conclusions: The correction equation described here is a simple method to correct for gadolinium interference on plasma selenium measurement using ICP-MS. This approach eliminates the need for specimen recollections, and improves patient care by reducing laboratory turnaround times and preventing delays in diagnosis/treatment.Keywords
This publication has 17 references indexed in Scilit:
- Inductively Coupled Plasma Mass Spectrometry: Introduction to Analytical Aspects2019
- 25 Years of Contrast-Enhanced MRI: Developments, Current Challenges and Future PerspectivesAdvances in Therapy, 2016
- Determination of selenium in serum in the presence of gadolinium with ICP-QQQ-MSThe Analyst, 2015
- Falsely elevated plasma selenium due to gadolinium contrast interference: a novel solution to a preanalytical problemAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2014
- Trace element analysis of human urine collected after administration of Gd-based MRI contrast agents: characterizing spectral interferences using inorganic mass spectrometryJournal of Analytical Atomic Spectrometry, 2013
- Interference of gadolinium on the measurement of selenium in human serum by inductively coupled plasma-quadrupole mass spectrometryAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2011
- Acute Selenium Toxicity Associated With a Dietary SupplementArchives of internal medicine (1960), 2010
- Regulation and function of selenoproteins in human diseaseBiochemical Journal, 2009
- Accidental death from acute selenium poisoningMedical Journal of Australia, 2006
- Gadolinium-DTPA as a contrast agent in MRI: initial clinical experience in 20 patientsAmerican Journal of Roentgenology, 1984