Behavior of a Polypyrrole Solid Contact pH-Selective Electrode Based on Tertiary Amine Ionophores Containing Different Alkyl Chain Lengths between Nitrogen and aPhenyl Group
Authors:
- Su-Jung YooSYSu-Jung Yoo
- Sung-Hoon KimSKSung-Hoon Kim
- Tae-Kee HongTHTae-Kee Hong
- Koo-Chun ChungKCKoo-Chun Chung
- 26 March 2003
- journal article
- Published by Springer Nature in Analytical Sciences
- Vol. 19 (3) , 357-360
Abstract
We compared the slopes of the EMF responses and the response range of a solid contact electrode based on a tribenzylamine ionophore with those based on a tris(2-phenylethyl)amine ionophore and a tris(3-phenylpropyl)amine ionophore. Their slopes of the EMF responses showed, respectively, 57.7, 58.1, 59.0 mV pH(-1) (at 20 +/- 0.2 degrees C). Also, the linear pH response dynamic ranges were 2.74-10.63 for tribenzylamine-based SCEs, pH 4.47-12.59 for tris(2-phenylethyl)amine, and pH 4.60-13.24 for tris(3-phenylpropyl)amine. When the slopes of the EMF responses and the response range were considered, as the length of the replaced alkyl chain increased between the nitrogen atom and the phenyl group in the ionophore, though the dynamic range and the slopes of the EMF responses of these electrodes did not show any particularly different results, their response ranges moved toward the alkali range (shifted to pH 14). When it was directly applied to cow's milk (in pH range of 6.0-8.5), we could obtain the same satisfying results. This electrode continuously contacted a Tris 7.50 buffered solution as well as cow's milk for one month without any loss of performance. Also, especially, these electrodes showed very good results, such as stabilization of the base potential as well as the best reproducibility and response time (< 10 s).Keywords
This publication has 13 references indexed in Scilit:
- Enhanced Electrochemical Performance of Poly(aniline) Solid-Contact pH Electrodes Based on AlkyldibenzylamineAnalytical Sciences, 2000
- Hydrogen ion selective electrode based on poly(1-aminoanthracene) filmAnalytica Chimica Acta, 1998
- All-Solid-State Potentiometric Sensors for Potassium and Sodium Based on Poly(pyrrole) Solid ContactMicrochemical Journal, 1997
- All solid-state hydrogen ion-selective electrode based on a conducting poly(pyrrole) solid contactThe Analyst, 1994
- Influence of basicity of dopant anions on the conductivity of polyanilineJournal of Electroanalytical Chemistry, 1992
- H+-selective electrodes based on neutral carriers: Specific features in behaviour and quantitative description of the electrode responseTalanta, 1990
- Design of neutral hydrogen ion carriers for solvent polymeric membrane electrodes of selected pH rangeAnalytical Chemistry, 1986
- Polyaniline, a novel conducting polymer. Morphology and chemistry of its oxidation and reduction in aqueous electrolytesJournal of the Chemical Society, Faraday Transactions 1: Physical Chemistry in Condensed Phases, 1986
- Coated-Wire Ion-Selective Electrodes1984
- A hydrogen ion-selective liquid-membrane electrode based on tri-n-dodecylamine as neutral carrierAnalytica Chimica Acta, 1981