Electronic-Textile 12-Lead Equivalent Diagnostic Electrocardiogram Based on the EASI Lead Placement
Authors:
- Meseret N. TeferraMTMeseret N. Teferra
- David A. HobbsDHDavid A. Hobbs
- Robyn A. ClarkRCRobyn A. Clark
- Karen J. ReynoldsKRKaren J. Reynolds
- 26 January 2022
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 22 (6) , 5994-6001
Abstract
Background: The standard 12-lead Holter ambulatory ECG monitor is complex in construction and uncomfortable for the wearer. A 12-lead equivalent ECG monitor was developed based on the EASI configuration with electronic-textile electrodes to meet these needs. Objective: In this study, we evaluated the accuracy and reliability of the EASI 12-lead equivalent ECG prototype. Results: The ECG hardware reproduced cardiac abnormalities from a ProSim 3.0 Vital signal simulator without significant deviation from a reference C.P. 200™ 12-Lead resting ECG. The vest and textile electrodes were subjected to 10 washing cycles. The joint connecting the snap fastener and the integrated wires were the weakest points, starting to fail after eight washing cycles; however, ten cycles of washing did not produce a noticeable loss of signal quality. The compression pressure for a stable-skin electrode interface of the vest was experimentally determined to be 15.92mmHg (minimum compression pressure at the ’I’ electrode position) and 29.45mmHg (maximum compression pressure at ’E’ electrode position). Conclusion: The textile-based ECG smart vest was non-inferior to standard 12-lead resting ECG and could be an alternative solution to monitor patients outside of a hospital setting, pending a larger scale trial with a clinical population.Keywords
This publication has 37 references indexed in Scilit:
- Effect of pressure and padding on motion artifact of textile electrodesBioMedical Engineering OnLine, 2013
- Evaluation of a textile-based wearable system for the electrocardiogram monitoring in cardiac patientsEP Europace, 2012
- Preliminary development of a wearable device for dynamic pressure measurement in garmentsProcedia Engineering, 2010
- Characterization of textile electrodes and conductors using standardized measurement setupsPhysiological Measurement, 2010
- The Birmingham rehabilitation uptake maximisation study (BRUM): a randomised controlled trial comparing home-based with centre-based cardiac rehabilitationHeart, 2008
- Application of PU-sealing into Cu/Ni electroless plated polyester fabrics for e-textilesFibers and Polymers, 2007
- Assessment of QT-measurement accuracy using the 12-lead electrocardiogram derived from EASI leadsJournal of Electrocardiology, 2007
- Diagnostic conclusions from the EASI-derived 12-lead electrocardiogram as compared with the standard 12-lead electrocardiogram in childrenAmerican Heart Journal, 2006
- Comparison of the Five-Electrode–Derived EASI Electrocardiogram to the Mason Likar Electrocardiogram in the Prehospital SettingThe American Journal of Cardiology, 2005
- Advantages of the mean absolute error (MAE) over the root mean square error (RMSE) in assessing average model performanceClimate Research, 2005