Combining Clinical LAFOV PET/CT with a Digital Twin Providing Motion-Free Ground Truth Reveals Quantitative Trade-offs in Respiratory Motion Correction
Preprint
Abstract
Purpose: Respiratory motion remains a major source of quantitative bias in PET and becomes increasingly relevant for high-sensitivity long axial field-of-view (LAFOV) PET/CT. Although numerous respiratory motion correction (MoCo) methods have been proposed, their quantitative accuracy cannot be established clinically because a patient-specific motion-free reference is fundamentally unavailable in vivo. This study combined clinical PET imaging with a digital twin, a realistic representation of both the PET/CT system and the patient, to objectively validate respiratory MoCo against a corresponding motion-free reference. Methods: Twenty patients (10 [¹⁸F]FDG with predominantly pulmonary lesions and 10 [¹⁸F]SiFA lin -TATE with predominantly hepatic lesions; total 135 lesions) were analyzed. The digital twin combined a validated LAFOV PET/CT simulation model with an anatomically realistic phantom containing 14 lung and liver lesions, two patient-derived respiratory patterns, and respiratory motion amplitudes of 2 and 3 cm, generating patient-like datasets with corresponding motion-free references. Data-driven and image-based MoCo were evaluated using lesion morphology, SUV mean , SUV max , and metabolic tumor volume (MTV). Results: In patients, data-driven MoCo produced larger SUV mean increases than image-based MoCo for liver (48.1±18.9% vs. 17.0±12.0%; pmax and larger MTV reductions. Simulation revealed marked motion-induced SUV mean underestimation before correction, particularly in liver (−31.2±6.8%) and lower lung (−15.5±13.9%). Relative to the motion-free reference, data-driven MoCo most accurately recovered hepatic uptake (4.3±11.7% vs. −10.0±9.2%; p=0.01) but overestimated pulmonary uptake (lower lung: 19.8±16.3% vs. −1.6±10.2%; p=0.02). SUV max showed the same regional behavior, whereas image-based MoCo yielded MTV estimates closer to the reference. Quantitative recovery was largely independent of respiratory pattern, while larger motion amplitudes mainly affected image-based MoCo. Conclusion: Combining clinical PET with a realistic digital twin and corresponding motion-free ground truth enabled objective validation of respiratory MoCo beyond conventional clinical evaluation. Larger correction-induced quantitative changes should not be equated with greater quantitative accuracy. Instead, MoCo performance was region- and metric-dependent, highlighting the value of ground-truth-based validation for developing and benchmarking respiratory motion correction and quantitative PET on LAFOV PET/CT systems.Keywords
Funding Information
- the Ministry for Science, Research and the Arts Baden-Württemberg
- Deutsche Forschungsgemeinschaft (INST 37/1145-1 FUGG)
- Germany’s Excellence Strategy (EXC 2180—390900677)
- the Medical Faculty of University of Tübingen
This publication has 32 references indexed in Scilit:
- Evaluation of a motion correction algorithm in lung cancer PET/CT: Phantom validation and patient studiesMedical Physics, 2025
- Monte Carlo simulation of two Siemens Biograph PET/CT system using GATE: Image quality performanceRadiation Physics and Chemistry, 2024
- Impact of patient motion on parametric PET imagingEuropean Journal of Nuclear Medicine and Molecular Imaging, 2024
- Design of an anthropomorphic PET phantom with elastic lungs and respiration modelingMedical Physics, 2021
- 3D Slicer as an image computing platform for the Quantitative Imaging NetworkMagnetic Resonance Imaging, 2012
- Evaluating FDG uptake changes between pre and post therapy respiratory gated PET scansRadiotherapy and Oncology, 2012
- The impact of respiratory motion on tumor quantification and delineation in static PET/CT imagingPhysics in Medicine & Biology, 2009
- Retrospective data-driven respiratory gating for PET/CTPhysics in Medicine & Biology, 2009
- Respiratory Motion in Positron Emission Tomography/Computed Tomography: A ReviewSeminars in Nuclear Medicine, 2008
- Organ Movement Reduction in PET/CT using Dual-Gated Listmode AcquisitionZeitschrift für Medizinische Physik, 2006