Bacterial contamination of PCA and epidural infusion devices

Authors:

Abstract
Summary: We prospectively audited the bacterial contamination of re‐useable analgesia infusion pumps. In a one‐month period, 112 samples from the handset and keypads of our analgesia infusion pumps were cultured for bacterial contamination. Forty‐five percent of handset swabs and 46% of keypad swabs grew bacteria; the commonest organism being coagulase‐negative staphylococcus. An additional cleaning step using 70% isopropyl alcohol wipes was introduced and the contamination rate was re‐audited in 100 samples. The contamination rate was reduced to 6% of handset swabs and 4% of keypad swabs. A high initial rate of bacterial contamination of re‐useable analgesia infusion pumps was significantly reduced by the implementation of a simple, additional cleaning procedure.