Using novel flow volume curve analysis to improve spirometry quality

Authors:

Abstract
Spirometry requires maximal forced exhalation, but no standard exists for quantifying patient effort. Submaximal effort can lead to falsely high FEV1 values (figure 1) through limiting dynamic compression of the airways, especially in lung conditions with loss of elasticity. erj;68/suppl_70/PA1746/F1 F1 F1 We investigated if similarity between flow volume curves could quantify spirometry effort and assessed its impact on FEV1 variability. Thirty eight COPD patients were assessed. Each curve from a spirometry session was exported and aligned on the volume axis to the curve with the highest PEF. The coefficient of determination (R2) was used to assess the similarity between curves, across the 1st litre exhaled, known as the effort dependent region of the flow volume curve. FEV1 and its range was compared between all curves and after discarding those with an R2 value 2 value 0.95 (mean[range] reduction in variability = -75 [-380,0] mls, p1 value (median[range] reduction= -10 [-130,0] mls, p<0.0001). Spirometry accuracy can be improved by quantifying the similarity between flow volume curves, excluding those with a coefficient of determination below 0.95. This is a simple way to automate spirometry quality, for both diagnostic purposes and for improving clinical trial data quality.