Determination of cetirizine and its impurities in bulk and tablet formulation using a validated capillary zone electrophoretic method
Authors:
- 25 April 2014
- journal article
- Published by Pleiades Publishing Ltd in Journal of Analytical Chemistry
- Vol. 69 (5) , 442-447
Abstract
A stability indicating capillary electrophoretic method for separation and determination of cetirizine dihydrochloride and its major impurities in bulk and a tablet dosage form was developed. The electrophoretic separation was performed in an uncoated fused-silica capillary (75 cm × 50 μm i.d.) using 75 mM sodium phosphate (pH 2.8) as background electrolyte, with an applied voltage of +25 kV at 25°C and UV detection at 230 nm. Fexofenadine was used as internal standard. The proposed method was found selective for determination of the main drug and its major impurities. The regression data obtained from the calibration plots indicated linear relationship (r 2 = 0.998) over the concentration range of 40–240 μg/mL of cetirizine. Repeatability and reproducibility of the method, assessed as intra-day and inter-day variation and expressed as RSD (%), were 1.3 and 2.6, respectively. Stress tests on cetirizine under acidic, basic, oxidative and heat incubating at 80°C conditions revealed that no major compound was formed under the applied conditions and the proposed CE method is applicable for stability studies on cetirizine. Then, the method was successfully applied to the determination of cetirizine in bulk and a tablet dosage form.Keywords
This publication has 17 references indexed in Scilit:
- Enantioseparation of cetirizine by sulfated‐β‐cyclodextrin‐mediated capillary electrophoresisJournal of Separation Science, 2008
- Development and validation of a rapid RP-HPLC method for the determination of cetirizine or fexofenadine with pseudoephedrine in binary pharmaceutical dosage formsJournal of Pharmaceutical and Biomedical Analysis, 2008
- Narrow-bore high performance liquid chromatographic method for the determination of cetirizine in human plasma using column switchingJournal of Pharmaceutical and Biomedical Analysis, 2005
- Determination of cetirizine dihydrochloride, related impurities and preservatives in oral solution and tablet dosage forms using HPLCJournal of Pharmaceutical and Biomedical Analysis, 2004
- Determination of cetirizine in human plasma by liquid chromatography-tandem mass spectrometryChromatographia, 2002
- Stability indicating HPTLC method for the simultaneous determination of pseudoephedrine and cetirizine in pharmaceutical formulationsJournal of Pharmaceutical and Biomedical Analysis, 2001
- Determination of cetirizine in human plasma by high-performance liquid chromatographyJournal of Chromatography B: Biomedical Sciences and Applications, 1999
- Spectrophotometric and high performance liquid chromatographic determination of cetirizine dihydrochloride in pharmaceutical tabletsJournal of Pharmaceutical and Biomedical Analysis, 1998
- Determination of cetirizine in human urine by high-performance liquid chromatographyJournal of Chromatography B: Biomedical Sciences and Applications, 1991
- Gas chromatographic method for the determination of cetirizine in plasmaJournal of Chromatography B: Biomedical Sciences and Applications, 1988