Accuracy of QuickScan Imaging Protocols of iCAT FLX CBCT in Assessment of Peri-implant Bone Defects

Authors:

Abstract
Aim of the study: The current study was conducted to determine the accuracy of QuickScan imaging protocols of iCAT Flx CBCT machine in detection and measuring crestal bone defects around dental implants.Materials and Methods: 20 simulated peri-implant crestal bone defects were created in bovine ribs bone models then were scanned by iCAT FLX CBCT machine with HD Scan 0.25 mm voxel, QuickScan 0.25mm voxel, and QuickScan + 0.3mm voxel sizes. All scans were imported in on-demand 3D App software and produced linear measurements were compared to true measurements done directly with digital caliper on bone models.Results: All imaging protocols of CBCT showed no statistical significance in detection of shallow peri-implant bone defects while the QuickScan + imaging protocol showed less accuracy in measuring the depth of defects.Conclusion: QuickScan + can be used as a low dose periodic follow up imaging utility for postoperative implant surgery for detection of any small bone changes but if more details are needed QuickScan technique is required.