Testing the accuracy of low-beam-energy electron-excited X-ray microanalysis with energy-dispersive spectrometry
Open Access
- 14 October 2024
- journal article
- research article
- Published by Springer Nature in Journal of Materials Science
- Vol. 59 (40) , 19088-19111
Abstract
The accuracy of electron-excited X-ray microanalysis with energy-dispersive spectrometry (EDS) has been tested in the low beam energy range, specifically at an incident beam energy of 5 keV, which is the lowest beam energy for which a useful characteristic X-ray peak can be excited for all elements of the periodic table, excepting H and He. Elemental analysis results are reported for certified reference materials (CRM), stoichiometric compounds, minerals, and metal alloys of independently known or measured composition which had microscopic homogeneity suitable for microanalysis. Two-hundred sixty-three concentration measurements for 39 elements in 113 materials were determined following the k-ratio protocol and using the EDS analytical software NIST DTSA-II. The accuracy of the results, as characterized by the relative deviation from expected value (RDEV) metric, was such that more than 98% of the results were found to be captured within a range of ±5% RDEV, while 82% of the results fell in the range -2% to 2% RDEV.Keywords
Funding Information
- National Institute of Standards and Technology
This publication has 16 references indexed in Scilit:
- Embracing Uncertainty: Modeling the Standard Uncertainty in Electron Probe Microanalysis—Part IMicroscopy and Microanalysis, 2020
- Using the EDS Clues: Peak Fitting Residual Spectrum and Analytical TotalMicroscopy and Microanalysis, 2019
- Scanning Electron Microscopy and X-Ray Microanalysis2018
- Theory and calculation of the atomic photoeffectRadiation Physics and Chemistry, 2016
- Electron Probe Microanalysis of Ni Silicides Using Ni-L X-Ray LinesMicroscopy and Microanalysis, 2016
- Electron-Excited X-Ray Microanalysis at Low Beam Energy: Almost Always an Adventure!Microscopy and Microanalysis, 2016
- Performing elemental microanalysis with high accuracy and high precision by scanning electron microscopy/silicon drift detector energy-dispersive X-ray spectrometry (SEM/SDD-EDS)Journal of Materials Science, 2014
- Low-Voltage Electron-Probe Microanalysis of Fe–Si Compounds Using Soft X-RaysMicroscopy and Microanalysis, 2013
- SEMantics for High Speed Automated Particle Analysis by SEM/EDXMicroscopy and Microanalysis, 2011
- Electron Channeling: A Problem for X-Ray Microanalysis in Materials ScienceMicroscopy and Microanalysis, 2009