A contribution to MRI safety testing related to gradient‐induced heating of medical devices
Open Access
- 28 March 2022
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 88 (2) , 930-944
Abstract
Purpose: To theoretically investigate the feasibility of a novel procedure for testing the MRI gradient‐induced heating of medical devices and translating the results into clinical practice.Methods: The concept of index of stress is introduced by decoupling the time waveform characteristics of the gradient field signals from the field spatial distribution within an MRI scanner. This index is also extended to consider the anisotropy of complex bulky metallic implants. Merits and drawbacks of the proposed index of stress are investigated through virtual experiments. In particular, the values of the index of stress evaluated for realistic orthopedic implants placed within an ASTM phantom are compared with accurate heating simulations performed with 2 anatomic body models (a man and a woman) implanted through a virtual surgery procedure.Results: The manipulation of the proposed index of stress allows to identify regions within the MRI bore where the implant could affect the safety of the examinations. Furthermore, the conducted analysis shows that the power dissipated into the implant by the induced eddy currents is a dosimetric quantity that estimates well the maximum temperature increase in the tissues surrounding the implant.Conclusion: The results support the adoption of an anisotropic index of stress to regulate the gradient‐induced heating of geometrically complex implants. They also pave the way for a laboratory characterization of the implants based on electrical measurements, rather than on thermal measurements. The next step will be to set up a standardized experimental procedure to evaluate the index of stress associated with an implant.Keywords
Funding Information
- Horizon 2020 Framework Programme
- European Commission
This publication has 34 references indexed in Scilit:
- Application and experimental validation of an integral method for simulation of gradient-induced eddy currents on conducting surfaces during magnetic resonance imagingPhysics in Medicine & Biology, 2013
- Heating of metallic rods induced by time‐varying gradient fields in MRIJournal of Magnetic Resonance Imaging, 2013
- Computational and experimental studies of an orthopedic implant: MRI‐related heating at 1.5‐T/64‐MHz and 3‐T/128‐MHzJournal of Magnetic Resonance Imaging, 2012
- Role of the lead structure in MRI‐induced heating: In vitro measurements on 30 commercial pacemaker/defibrillator leadsMagnetic Resonance in Medicine, 2011
- MRI Induced Heating of Artificial Bone ImplantsJournal of Electromagnetic Waves and Applications, 2009
- Spatial distribution of RF‐induced E‐fields and implant heating in MRIMagnetic Resonance in Medicine, 2008
- Food and Drug Administration PerspectiveCirculation, 2006
- Current Clinical Issues for MRI Scanning of Pacemaker and Defibrillator PatientsPacing and Clinical Electrophysiology, 2005
- Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heatingJournal of Magnetic Resonance Imaging, 2004
- Choosing a Point from the Surface of a SphereThe Annals of Mathematical Statistics, 1972