Heat loss during carbon dioxide insufflation: comparison of a nebulization based humidification device with a humidification and heating system
Authors:
- 22 June 2012
- journal article
- research article
- Published by Springer Nature in Surgical Endoscopy
- Vol. 26 (12) , 3622-3625
Abstract
This study compared the heat loss observed with the use of MR860 AEA Humidifier™ system (Fisher & Paykel Healthcare, New Zealand), which humidifies and heats the insufflated CO2, and the use of the AeronebPro™ device (Aerogen, Ireland), which humidifies but does not heat the insufflated CO2. With institutional approval, 16 experiments were conducted in 4 pigs. Each animal, acting as its own control, was studied at 8-day intervals in randomized sequence with the following four conditions: (1) control (C) no pneumoperitoneum; (2) standard (S) insufflation with nonhumidified, nonheated CO2; (3) Aeroneb™ (A): insufflation with humidified, nonheated CO2; and (4) MR860 AEA humidifier™ (MR): insufflation with humidified and heated CO2. The measured heat loss after 720L CO2 insufflation during the 4 h was 1.03 ± 0.75 °C (mean ± SEM) in group C; 3.63 ± 0.31 °C in group S; 3.03 ± 0.39 °C in group A; and 1.98 ± 0.09 °C in group MR. The ANOVA showed a significant difference with time (p = 0.0001) and with the insufflation technique (p = 0.024). Heat loss in group C was less than in group S after 60 min (p = 0.03), less than in group A after 70 min (p = 0.03), and less than in group MR after 150 min (p = 0.03). The heat loss in group MR was less than in group S after 50 min (p = 0.04) and less than in group A after 70 min (p = 0.02). After 160 min, the heat loss in group S was greater than in group A (p = 0.03). As far as heat loss is concerned, for laparoscopic procedures of less than 60 min, there is no benefit of using any humidification with or without heating. However, for procedures greater than 60 min, use of heating along with humidification, is superior.Keywords
This publication has 20 references indexed in Scilit:
- Changes in Core Temperature During Peritoneal Insufflation: Comparison of Two CO2 Humidification Devices in PigsJournal of Surgical Research, 2011
- Heated CO2with or without humidification for minimally invasive abdominal surgery2011
- Independent testing of the Fisher & Paykel Healthcare MR860 Laparoscopic Humidification SystemMinimally Invasive Therapy & Allied Technologies, 2010
- Thermoregulatory defense mechanismsCritical Care Medicine, 2009
- An Evaluation of Gas Humidifying Devices as a Means of Intraperitoneal Local Anesthetic Administration for Laparoscopic SurgeryAnesthesia & Analgesia, 2008
- Cold nebulization used to prevent heat loss during laparoscopic surgery: an experimental study in pigsSurgical Endoscopy, 2008
- The Effect of Heated Humidified Carbon Dioxide on Postoperative Pain, Core Temperature, and Recovery Times in Patients Having Laparoscopic Surgery: A Randomized Controlled TrialJournal of Minimally Invasive Gynecology, 2008
- Prospective randomized trial of heated humidified versus cold dry carbon dioxide insufflation during laparoscopic gastric bypassSurgery for Obesity and Related Diseases, 2006
- Perioperative Heat BalanceAnesthesiology, 2000
- Perioperative Normothermia to Reduce the Incidence of Surgical-Wound Infection and Shorten HospitalizationThe New England journal of medicine, 1996