Evaluation of the Elekta Agility MLC performance using high‐resolution log files
- 31 January 2019
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 46 (3) , 1397-1407
Abstract
Purpose: With the advent of volumetric modulated arc therapy (VMAT) and intensity‐modulated radiation therapy (IMRT) treatment techniques, the requirement for more elaborate approaches in reviewing linac components’ integrity has become even more stringent. A possible solution to this challenge is to employ the usage of log files generated during treatment. The log files generated by the new generation of Elekta linacs record events at a higher frequency (25 Hz) than their predecessors, which allows for retrospective analysis and identification of subtle changes and provides another means of quality assurance. The ability to track machine components based on log files for each treatment can allow for constant monitoring of fraction consistency in addition to machine reliability. Using Elekta Agility log files, a set of tests were developed to evaluate the reliability and robustness of the multileaf collimators (MLCs).Methods: To evaluate Elekta log file utilization for linac MLC QA effectiveness, five MLC test patterns were constructed to review the effects of leaf velocity and acceleration on positional accuracy, including gravitational effects for the Elekta MLC system. Each test was run five times in a particular setting to obtain reproducibility data and statistical averages. This study was performed on two identical Versa HD machines, each delivering a full set of test plans with all possible variations. Plans were delivered using Elekta's iCOMcat software and recorded log files were extracted. Log files were reformatted for readability and automatically analyzed in Matlab®.Results: The Elekta Agility MLC system was shown to be capable of obtaining speeds within the range of 5–35 mm/s. MLC step and shoot tests have demonstrated the MLC system's capability of having positional repeatability, averaging 0.03‐ and 0.08‐mm offsets with and without gravitational effects, respectively. The IMRT‐specific tests have shown that gravitational effects are negligible with all positional tests averaging 0.5‐mm offsets. The largest speed root‐mean‐square error (RMSE) for the MLC system was found at the maximum speed of 35 mm/s with an average error of 0.8 mm. For slower speeds, the value was found to be much lower.Conclusion: Utilizing log files has demonstrated the feasibility for higher precision of MLC motions to be reviewed, based on the performance tests that were instituted. Log files provide insight on the effects of friction, acceleration, and gravity, with MU's delivered that previously could not be reviewed in such detail. Based on our results, log file‐based QA has enhanced our ability to review performance, functionality, and perform QA on Elekta's MLC system.Keywords
Funding Information
- Cancer Prevention and Research Institute of Texas
- Elekta
This publication has 12 references indexed in Scilit:
- Investigation of error detection capabilities of phantom, EPID and MLC log file based IMRT QA methodsJournal of Applied Clinical Medical Physics, 2017
- Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveriesPhysics in Medicine & Biology, 2014
- A comprehensive tool to analyse dynamic log files from an Elekta-Synergy acceleratorJournal of Physics: Conference Series, 2014
- Catching errors with patient-specific pretreatment machine log file analysisPractical Radiation Oncology, 2013
- The inter‐ and intrafraction reproducibilities of three common IMRT delivery techniquesMedical Physics, 2010
- Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log filesPhysics in Medicine & Biology, 2010
- Physical Characterization of a New Concept Design of an Elekta Radiation Head with Integrated 160-leaf Multi-leaf CollimatorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Development and evaluation of an efficient approach to volumetric arc therapy planningMedical Physics, 2009
- An extensive log‐file analysis of step‐and‐shoot intensity modulated radiation therapy segment delivery errorsMedical Physics, 2004