Differential input amplifiers are commonly used in the measurement of biopotential signals. This is because of their ability to reject common-mode signals, for example power-line interference. Rejection is further enhanced by referencing the subject to the instrumentation system. One such referencing technique uses a third subject electrode connected either to circuit common or to ground. However, these techniques are rarely explained or quantified in the literature. Four recording configurations are examined, isolated and non-isolated amplifiers with two and with three electrodes, and existing models are extended. From a brief assessment of power-line interference sources, it is concluded that only displacement currents induced into the leads and the body need be considered. The effect of these on the four configurations is analysed using SPICE. Results show that interference from displacement currents induced into the leads and into the body is dependent upon the recording configuration used and is significantly lower in three-electrode than in two-electrode amplifiers. Isolation also reduces levels of interference. It is concluded that the choice of referencing may affect interference levels on the biopotential signal significantly. Experimental and further analytical work is planned to confirm this.Keywords
This publication has 15 references indexed in Scilit:
- Skin impedance from 1 Hz to 1 MHzIEEE Transactions on Biomedical Engineering, 1988
- Interference-rejection characteristics of biopotential amplifiers: a comparative analysisIEEE Transactions on Biomedical Engineering, 1988
- Reducing Motion Artifacts and Interference in Biopotential RecordingIEEE Transactions on Biomedical Engineering, 1984
- Mains interference in ECG recordingJournal of Medical Engineering & Technology, 1984
- Driven-right-leg circuit designIEEE Transactions on Biomedical Engineering, 1983
- Ground-Free ECG Recording with Two ElectrodesIEEE Transactions on Biomedical Engineering, 1980
- How to prevent 60-Hz interference.1979
- Miniature skin-mounted preamplifier for measurement of surface electromyographic potentialsMedical & Biological Engineering & Computing, 1977
- An analysis of the use of active electrodes in electroencephalogram ambulatory monitoring.1976
- 60-Hz Interference in ElectrocardiographyIEEE Transactions on Biomedical Engineering, 1973