Design and characterization of a versatile reference instrument for rapid, reproducible specular gloss measurements
Authors:
- 1 August 2005
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 44 (22) , 4631-4638
Abstract
A reference goniospectrophotometer has been developed at the National Research Council of Canada (NRC) for providing high-accuracy traceable measurements of specular gloss at several standard geometries, including 75° for paper samples, haze and absence-of-bloom gloss, and color appearance of gonioapparent materials. This is to the authors’ knowledge the first reported reference instrument that has this level of versatility for rapidly characterizing the total visual appearance properties of a wide variety of materials and applications. This instrument also replaces the NRC glossmeter that has been providing primary level specular gloss measurements in accordance with International Organization for Standardization and American Society for Testing and Materials standards for measurements of paint and ceramic materials at geometries of 20°, 60°, and 85°. The new instrument has been fully characterized for sources of error and compared with the NRC glossmeter. Its measurement reproducibility of 0.02 gloss unit is a factor-of-5 improvement, and its overall estimated expanded (k = 2) uncertainty is 0.3 gloss unit at all three standard geometries.Keywords
This publication has 8 references indexed in Scilit:
- Specular gloss scales comparison between the national institute of standards and technology and the national research council of canadaJournal of Coatings Technology, 2003
- NIST reference goniophotometer for specular gloss measurementsJournal of Coatings Technology, 2001
- D65 simulation with a xenon arcApplied Optics, 1996
- Depolarization measurements of an integrating sphereApplied Optics, 1995
- The Calibration of Gloss Reference StandardsMetrologia, 1980
- A Reference Instrument for 20°, 60° and 85° Gloss MeasurementsMetrologia, 1980
- Polarization effects in gloss measurementsApplied Optics, 1979
- Design of Glass Filter Combinations for PhotometersApplied Optics, 1969