Development of 3-D detector system for positron CT
Authors:
- 1 February 1988
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 35 (1) , 717-720
Abstract
A detector composed of a bundle of BGO (bismuth germanate) pillars and position-sensitive photomultiplier tubes is proposed. This detector has the ability to provide 3-D position information on gamma -ray absorption in the scintillator. It is expected that the use of this detector for positron-emission tomography (PET) will result in high spatial resolution and sensitivity. In particular, uniform and high spatial resolution over the entire field of view can be obtained because of the detector's ability to detect the depth of interaction of the gamma -rays in the scintillator. The fundamental characteristics were investigated by both experiment and simulation, and the results are discussed.<>Keywords
This publication has 13 references indexed in Scilit:
- Depth-of-interaction determination in NaI(Tl) and BGO scintillation crystals using a temperature gradientNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1987
- Design of an efficient position sensitive gamma ray detector for nuclear medicinePhysics in Medicine & Biology, 1986
- Design of a Mosaic BGO Detector System for Positron CTIEEE Transactions on Nuclear Science, 1986
- Position Sensitive Photomultiplier Tubes for Scintillation ImagingIEEE Transactions on Nuclear Science, 1986
- Designing a Stratified Detection System for PET CamerasIEEE Transactions on Nuclear Science, 1986
- A Multicrystal Two Dimensional BGO Detector System for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1986
- A Modular Scintillation Camera for Use in Nuclear MedicineIEEE Transactions on Nuclear Science, 1984
- Initial Characterization of a Bgo-Photodiode Detector for High Resolution Positron Emission TomographyIEEE Transactions on Nuclear Science, 1984
- A large position-sensitive neutron detectorNuclear Instruments and Methods in Physics Research, 1981
- PETT IV: Design Analysis and Performance CharacteristicsIEEE Transactions on Nuclear Science, 1978