Finite element analysis of low-profile reconstruction plates for atrophic mandibles: a comparison of novel 3D grid and conventional plate designs
Authors:
- Raphael Capelli GuerraRaphael Capelli Guerra
- Bianca de Fátima Borim PulinoBPBianca de Fátima Borim Pulino
- Vitor Floriano Salomão JúniorVJVitor Floriano Salomão Júnior
- Rodrigo dos Santos PereiraRPRodrigo dos Santos Pereira
- Florian M. ThieringerFlorian M. Thieringer
- Roberto SaccoRoberto Sacco
- Robert SaderRSRobert Sader
- Eduardo Hochuli VieiraEVEduardo Hochuli Vieira
- 29 July 2023
- journal article
- research article
- Published by Springer Nature in Oral and Maxillofacial Surgery
- Vol. 28 (2) , 595-603
Abstract
Objectives This study aimed to compare the reconstruction with type 2.4 and three-dimensional (3D) grid plates relating the areas of weakness to anatomical regions of force incidence in atrophic mandibles with and without segmentation using finite element analysis (FEA). Materials and methods Strength and force quality in the mandible were also determined. The atrophic mandible models with and without segmentation were divided into four groups, and the behavior of each plate was analyzed using finite element analysis. In the atrophic mandible without segmentation using the type 2.4 reconstruction plate, the highest stress concentration in the plate was observed in the posterior region of the mandibular body and in the grid type 3D reconstruction plate was observed in the mandibular angle. Results In the segmented atrophic mandible with the 2.4 reconstruction plate, higher stress concentration was observed in the mandibular segment defect. Analysis of the 3D grid-like reconstruction plate revealed that the geometry of the plate conferred greater stiffness to the assembly, as the most significant stress concentration was observed at the mandibular angle. Conclusion The width of the plate design may influence the strength, not the thickness.Keywords
All Related Versions
This publication has 28 references indexed in Scilit:
- Mandibular Resection and Reconstruction in the Management of Extensive AmeloblastomaJournal of Oral and Maxillofacial Surgery, 2013
- The management of unstable oblique infected mandibular fractures with a 2.3 mm mandibular osteosynthesis reconstruction bone plateJournal of Cranio-Maxillofacial Surgery, 2011
- Review of biomechanical models used in studying the biomechanics of reconstructed mandiblesInternational Journal of Oral and Maxillofacial Surgery, 2011
- Contemporary reconstruction of the mandibleOral Oncology, 2010
- Evaluation of different miniplates in fixation of fractured human mandible with the finite element methodOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2007
- Adaquate fixation of plates for stability during mandibular reconstructionJournal of Cranio-Maxillofacial Surgery, 2006
- Analysis of mechanical stress in reconstruction plates for bridging mandibular angle defectsJournal of Cranio-Maxillofacial Surgery, 2006
- Treatment of Atrophic Mandibular Fractures Based on the Degree of Atrophy—Experience With Different Plating Systems: A Retrospective StudyJournal of Oral and Maxillofacial Surgery, 2006
- Comparison of different fixation methods following sagittal split ramus osteotomies using three-dimensional finite elements analysisInternational Journal of Oral and Maxillofacial Surgery, 2005
- Experimentelle Untersuchung zur biomechanischen Stabilität unterschiedlicher UnterkieferosteosynthesenMund-,Kiefer- und Gesichtschirurgie, 2003