Long‐term performance and lead failure analysis of the Durata defibrillation lead compared to its previous model, the recalled Riata defibrillation lead
Authors:
- 22 July 2019
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 30 (10) , 2012-2019
Abstract
Introduction: Data on long‐term durability of St Jude Medical Durata defibrillation leads compared to its previous model, the St Jude Medical Riata leads in clinical practice are missing. Aim of the study was to analyze the long‐term performance of the Durata defibrillation leads compared to the Riata leads in clinical practice.Methods and Results: A total of 1407 consecutive patients of a prospective single‐centre implantable cardioverter defibrillator (ICD)‐registry were analyzed who underwent ICD implantation with a Durata (n = 913) or Riata (n = 494) ICD lead between 2002 and 2017. Most of the leads were implanted via a subclavian vein access. The estimated lead defect rates after 5 and 10 years were not different between the Durata (11% and 36%) and Riata leads (13% and 38%). Among Durata leads single coil and DF‐4 connector ICD leads had a lower incidence of lead failure. Major causes of lead failure were compression of the lead in the clavicular region, generator to lead friction and distal fatigue fracture whereas lead defect due to externalization was a rare cause of lead defect in Riata leads (3%).Conclusion: Among ICD leads implanted via the subclavian vein access the lead defect rate of Durata leads after 10 years is similar to that of Riata leads. Single coil and DF‐4 ICD leads are associated with a lower lead failure rate. Mechanical stress represents a major cause of lead failure mechanism whereas externalization might only play a minor role in clinical practice.Keywords
This publication has 20 references indexed in Scilit:
- Prospective long-term evaluation of Optim-insulated (Riata ST Optim and Durata) implantable cardioverter-defibrillator leadsHeart Rhythm, 2014
- Prevalence of externalized conductors in Riata and Riata ST silicone leads: Results from the prospective, multicenter Riata Lead Evaluation StudyHeart Rhythm, 2013
- Early fatigue fractures in the IS-1 connector leg of a small-diameter ICD lead: Value of returned product analysis for improving device safetyHeart Rhythm, 2013
- Mechanisms, predictors, and trends of electrical failure of Riata leadsHeart Rhythm, 2013
- Prevalence and Presentation of Externalized Conductors and Electrical Abnormalities in Riata Defibrillator Leads After Fluoroscopic ScreeningCirculation: Arrhythmia and Electrophysiology, 2012
- Failure of a novel silicone-polyurethane copolymer (Optim ) to prevent implantable cardioverter-defibrillator lead insulation abrasionsEP Europace, 2012
- Clinical Predictors of Fidelis Lead FailureCirculation, 2012
- Insulation Defects of Thin High-Voltage ICD Leads:Journal of Cardiovascular Electrophysiology, 2011
- Sensing Failure Associated with the Medtronic Sprint Fidelis Defibrillator LeadJournal of Cardiovascular Electrophysiology, 2007
- A new insulation material for cardiac leads with potential for improved performanceHeart Rhythm, 2005