A New Stand-Alone Cervical Anterior Interbody Fusion Device
Authors:
- Matti ScholzMSMatti Scholz
- Phillip M. ReyesPRPhillip M. Reyes
- Philipp SchleicherPSPhilipp Schleicher
- Anna G. U. SawaASAnna G. U. Sawa
- Seungwon BaekSBSeungwon Baek
- Frank KandzioraFKFrank Kandziora
- Frederick F. MarcianoFMFrederick F. Marciano
- Neil R. CrawfordNCNeil R. Crawford
- 1 January 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Spine
- Vol. 34 (2) , 156-160
Abstract
A new anchored spacer-a low-profile cervical interbody fusion cage with integrated anterior fixation-was compared biomechanically to established anterior cervical devices. To evaluate the fixation properties of the new stand-alone device and compare these properties with established fixation methods. The hypothesis is that the new device will provide stability comparable to that provided by an anterior cervical cage when supplemented with an anterior plate. It is accepted that the use of anterior cervical plating increases the chance of achieving a solid fusion. However, its use may be associated with an increase in operation time and a higher postoperative morbidity caused by a larger anterior approach and disruption of the anterior musculature. This dilemma has led to the development of a new, low profile stand-alone cervical anterior cage device with integrated screw fixation. Twenty-four human cadaveric C4-C7 cervical spines were loaded nondestructively with pure moments in a nonconstraining testing apparatus to induce flexion, extension, lateral bending, and axial rotation while angular motion was measured optoelectronically. The specimens were tested: 1. Intact (N = 24). 2. After discectomy and anterior stabilization. a. Interbody cage + locking plate (N = 8). b. Interbody cage + dynamic plate (N = 8). c. Anchored spacer (N = 8). 3. After ventral plate removal of group 2a and 2b (N = 16). All fixation techniques decreased range of motion (ROM) and lax zone (LZ) (P < 0.05) in all test modes compared with the intact motion segment and cage-only group. There were no significant differences between the anchored spacer and cage + locking plate or cage + dynamic plate. The anchored spacer provided a similar biomechanical stability to that of the established anterior fusion technique using an anterior plate plus cage and has a potentially lower perioperative and postoperative morbidity. These results support progression to clinical trials using the cervical anchored spacer as a stand-alone implant.Keywords
This publication has 12 references indexed in Scilit:
- Esophageal perforation from anterior cervical screw migrationSurgical Neurology, 2007
- A New Stand-Alone Anterior Lumbar Interbody Fusion Device: Biomechanical Comparison with Established Fixation TechniquesSpine, 2005
- Influence of Anterior Cervical Plate Design on DysphagiaJournal of Spinal Disorders & Techniques, 2005
- Biomechanical Comparison of Cervical Spine Interbody Fusion CagesSpine, 2001
- A new technique for determining 3-D joint angles: the tilt/twist methodClinical Biomechanics, 1999
- The Spinal Lax Zone and Neutral ZoneJournal of Spinal Disorders, 1998
- The Significance of Hardware Failure in Anterior Cervical Plate FixationSpine, 1998
- Construction of Local Vertebral Coordinate Systems Using a Digitizing ProbeSpine, 1997
- An Apparatus for Applying Pure Nonconstraining Moments to Spine Segments In VitroSpine, 1995
- Anterior Plate Stabilization for Fracture-dislocations of the Lower Cervical Spine1980