Angiography‐guided mid/high septal implantation of ventricular leads in patients with congenital heart disease
Open Access
- 17 September 2021
- journal article
- research article
- Published by Wiley in Journal of Arrhythmia
- Vol. 37 (6) , 1512-1521
Abstract
Background Conduction system pacing prevents pacing-induced cardiomyopathy, but it can be challenging to perform in patients with congenital heart disease (CHD), and mid/high septal lead implantation is an alternative. This study aimed to assess intraprocedural angiography's utility as a guide for mid/high-septal lead implantation in CHD patients. Methods The study subjects were CHD patients with Class I/IIa indications for permanent pacemaker implantation. To guide septal lead implantation, we performed an intraprocedural right ventricular angiogram in anteroposterior, 40° left anterior oblique, and 30° right anterior oblique. The primary endpoint was the lead tip in the mid/high septum on computed tomography (CT). The secondary endpoints were complications and systemic ventricular function on follow-up. Results From January 2008 to December 2018, we enrolled 27 patients (mean age: 30 ± 20 years; M:F 17:10) with CHD (unoperated: 20, operated: 7). The mean paced QRS duration was 131.7 ± 5.8 ms, and CT done in 22/27 patients confirmed the lead tip in the mid-septum in 16, high septum in 5, and apical septum in 1 patient. There were no procedural complications, and during a mean follow-up of 58 ± 35.2 months, there was no significant change in the systemic ventricular ejection fraction (56.4 ± 8.3% vs 53.9 + 5.9%, P = .08). Two patients with Eisenmenger syndrome died because of refractory heart failure. Conclusions Intraprocedural angiography is safe and useful to guide mid/high-septal lead implantation in CHD patients. Mid/high septal lead position preserves systemic ventricular function in patients with CHD during medium-term follow-up.Keywords
This publication has 24 references indexed in Scilit:
- Cardiac pacing in paediatric patients with congenital heart defects: transvenous or epicardial?EP Europace, 2013
- Impact of temporary right ventricular pacing from different sites on echocardiographic indices of cardiac functionEP Europace, 2011
- Predictors of left ventricular remodelling and failure in right ventricular pacing in the youngEuropean Heart Journal, 2009
- ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm AbnormalitiesHeart Rhythm, 2008
- Technical Aspects of Pacing in Adult and Pediatric Congenital Heart DiseasePacing and Clinical Electrophysiology, 2008
- Tricks of the Trade2007
- Know the Anatomy2007
- Detrimental Ventricular Remodeling in Patients With Congenital Complete Heart Block and Chronic Right Ventricular Apical PacingCirculation, 2004
- Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart diseaseHeart Rhythm, 2004
- Selective Site Pacing:Pacing and Clinical Electrophysiology, 2004