Abstract
I have demonstrated the advantage of the phase locked oscillator (PLO) over the conventional direct resonance oscillator (DRO) in noncontact mode atomic force microscopy (AFM) under ultrahigh vacuum. Direct comparison between PLO and DRO has been made in terms of background noise level, temporal response, and stability. Compared to the DRO method without phase coherence, the experimental results show that the PLO method is more effective in reducing the noise level and enhancing the stability over all force regimes in UHV noncontact AFM. The noise reduction and stability enhancement in PLO indicate the important role of the phase coherent effect in improving the capability of noncontact imaging in UHV.Keywords
This publication has 17 references indexed in Scilit:
- Phase variation experiments in non-contact dynamic force microscopy using phase locked loop techniquesApplied Surface Science, 1999
- Dynamic force microscopy by means of the phase-controlled oscillator methodJournal of Applied Physics, 1997
- Magnetic dissipation force microscopyApplied Physics Letters, 1997
- A phase-locked shear-force microscope for distance regulation in near-field optical microscopyApplied Physics Letters, 1997
- Atomic resolution in dynamic force microscopy across steps on Si(1 1 1)7×7Zeitschrift für Physik B Condensed Matter, 1996
- Defect Motion on an InP(110) Surface Observed with Noncontact Atomic Force MicroscopyScience, 1995
- Atomic Resolution of the Silicon (111)-(7×7) Surface by Atomic Force MicroscopyScience, 1995
- Observation of 7×7 Reconstructed Structure on the Silicon (111) Surface using Ultrahigh Vacuum Noncontact Atomic Force MicroscopyJapanese Journal of Applied Physics, 1995
- Interaction force detection in scanning probe microscopy: Methods and applicationsJournal of Applied Physics, 1992
- Frequency modulation detection using high-Q cantilevers for enhanced force microscope sensitivityJournal of Applied Physics, 1991