Impact of Disinfection and Sterilization on 3D-Printing Resin Performance for Surgical Guides in Cardiac Ablation Surgery
Authors:
- Rani KronenbergerRKRani Kronenberger
- Rawan KazmaRKRawan Kazma
- Alireza AmirabadiAAAlireza Amirabadi
- Leire Viana UribeLULeire Viana Uribe
- Giacomo TaleviGTGiacomo Talevi
- Görkem Eylül KayaGKGörkem Eylül Kaya
- Niko Van Den BrandeNBNiko Van Den Brande
- Ramak Hossein AbadiRARamak Hossein Abadi
- Kalliopi-Artemi KalteremidouKKKalliopi-Artemi Kalteremidou
- Danny Van HemelrijckDHDanny Van Hemelrijck
- Kitty BaertKBKitty Baert
- Tom HauffmanTHTom Hauffman
Open Access
- 28 August 2025
- journal article
- research article
- Published by MDPI AG in Bioengineering
- Vol. 12 (9) , 924
Abstract
Patient-tailored, 3D-printed surgical guides offer significant potential to improve precision and therapeutic efficacy in cardiac ablation surgery. However, reliable post-sterilization material performance presents a critical yet underexplored barrier to clinical adoption. This study investigates how disinfection and sterilization impact the mechanical and thermal properties of photopolymer resins. Specimens from two 3D-printing resins (Bioflex A80 MB™, 3Dresyns; MED625FLX™, Stratasys) were treated with four combinations of disinfection techniques (low-temperature manual cleaning; high-temperature machine washing) and sterilization techniques (H2O2 vs. autoclaving). We assessed post-sterilization properties by mechanical (material integrity, bending tests), thermal (differential scanning calorimetry, thermogravimetric analysis), and viscoelastic (dynamic mechanical analysis) studies. Statistical analysis was performed using one-way ANOVA with Bonferroni post hoc tests (α = 0.05). From this preliminary study, we conclude that MED625FLX maintains integrity and flexibility across all tested disinfection and sterilization methods. Bioflex A80 MB is only suitable for low-temperature disinfection–sterilization, as high-temperature treatments cause surface cracking. Neither resin is appropriate for cryogenic conditions due to the risk of brittleness. Further research into post-sterilization properties is essential to ensure the safety and clinical reliability of these materials in cardiac procedures.Keywords
This publication has 32 references indexed in Scilit:
- Assessing Leachable Cytotoxicity of 3D-Printed Polymers and Facile Detoxification Methods3D Printing and Additive Manufacturing, 2023
- 3D mapping challenges in hybrid video-assisted thoracoscopic surgical ablation of Brugada syndromeInterdisciplinary CardioVascular and Thoracic Surgery, 2023
- Sterilization of Polymeric Implants: Challenges and OpportunitiesACS Applied Bio Materials, 2022
- Biocompatible 3D printing resins for medical applications: A review of marketed intended use, biocompatibility certification, and post-processing guidanceAnnals of 3D Printed Medicine, 2022
- Sterilization of bedside 3D-printed devices for use in the operating roomAnnals of 3D Printed Medicine, 2022
- Dental resins used in 3D printing technologies release ovo-toxic leachatesChemosphere, 2021
- Bacteriological and mechanical impact of the Sterrad sterilization method on personalized 3D printed guides for mandibular reconstructionScientific Reports, 2021
- Effect of Steam Heat Sterilization on the Accuracy of 3D Printed Surgical GuidesImplant Dentistry, 2019
- Sterilization of implantable polymer-based medical devices: A reviewInternational Journal of Pharmaceutics, 2018
- Methods for Verifying Medical Device Cleanliness2015