High efficiency extreme ultraviolet overview spectrometer: Construction and laboratory testing
Authors:
- 22 September 2006
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 77 (10) , 10F305
Abstract
The new high efficiency extreme ultraviolet overview spectrometer system (HEXOS) has been developed to study impurity concentrations and impurity transport properties in the plasma of the stellarator W7-X. The HEXOS system consists of four different grating based spectrometers, which provide large etendue and good spectral resolution over a broad wavelength range ( , divided into four subsections with some overlapping). The mechanical arrangement as two double spectrometers allows for a compact installation geometry on W7-X. Laboratory testing and wavelength and intensity calibrations have been performed using a dc hollow cathode discharge and a pinch discharge .
Keywords
This publication has 12 references indexed in Scilit:
- Design of a high–efficiency extreme ultraviolet overview spectrometer system for plasma impurity studies on the stellarator experiment Wendelstein 7-XReview of Scientific Instruments, 2004
- Detector system with high time resolution for the continuous measurement of spectra in the vacuum ultraviolet wavelength rangeReview of Scientific Instruments, 2004
- High current hollow cathode as a radiometric transfer standard source for the extreme vacuum ultravioletApplied Optics, 1988
- High-resolution duo-multichannel soft x-ray spectrometer for tokamak plasma diagnosticsReview of Scientific Instruments, 1987
- SPRED spectrograph upgrade: High-resolution grating and improved absolute calibrationsReview of Scientific Instruments, 1986
- Impurity and radiation studies during the JET Ohmic Heating PhaseNuclear Fusion, 1986
- Impurities in tokamaksNuclear Fusion, 1984
- Grazing incidence time-resolving spectrograph for magnetic fusion plasma diagnosticsReview of Scientific Instruments, 1984
- Mechanically ruled aberration-corrected concave gratings for a flat-field grazing-incidence spectrographApplied Optics, 1983
- Multichannel grazing-incidence spectrometer for plasma impurity diagnosis: SPREDApplied Optics, 1982