Stretchable PPG sensor with light polarization for physical activity–permissible monitoring
Open Access
- 15 April 2022
- journal article
- research article
- Published by American Association for the Advancement of Science (AAAS) in Science Advances
- Vol. 8 (15) , eabm3622
Abstract
Skin-attachable sensors, which represent the ultimate form of wearable electronic devices that ensure conformal contact with skin, suffer from motion artifact limitations owing to relative changes in position between the sensor and skin during physical activities. In this study, a polarization-selective structure of a skin-conformable photoplethysmographic (PPG) sensor was developed to decrease the amount of scattered light from the epidermis, which is the main cause of motion artifacts. The motion artifacts were suppressed more than 10-fold in comparison with those of rigid sensors. The developed sensor—with two orthogonal polarizers—facilitated successful PPG signal monitoring during wrist angle movements corresponding to high levels of physical activity, enabling continuous monitoring of daily activities, even while exercising for personal health care.Keywords
This publication has 33 references indexed in Scilit:
- Highly Skin‐Conformal Microhairy Sensor for Pulse Signal AmplificationAdvanced Materials, 2014
- Peripheral Perfusion Index as an Early Predictor for Central Hypovolemia in Awake Healthy VolunteersAnesthesia & Analgesia, 2013
- Reducing motion artifacts in photoplethysmograms by using relative sensor motion: phantom studyJournal of Biomedical Optics, 2012
- Optical properties of human skinJournal of Biomedical Optics, 2012
- Photoplethysmography: Beyond the Calculation of Arterial Oxygen Saturation and Heart RateAnesthesia & Analgesia, 2007
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- Random Walk of Polarized Light in Turbid MediaPhysical Review Letters, 2005
- Fluorescence polarization analysis of protein-DNA and protein-protein interactions.Molecular Endocrinology, 1996
- Depolarization of multiply scattered waves by spherical diffusers: Influence of the size parameterPhysical Review E, 1994
- The role of exercise testing in evaluation of arrhythmiasThe American Journal of Cardiology, 1988