Clinical Interpretation of Peripheral Pulse Oximeters Labeled “Not for Medical Use”
Open Access
- 12 November 2018
- journal article
- research article
- Published by Annals of Family Medicine in The Annals of Family Medicine
- Vol. 16 (6) , 552-554
Abstract
The purpose of our study was to clarify limitations of off-label use for low cost nonmedical use (NMU) pulse oximeters by primary care providers. These devices are widely marketed over the Internet and in drugstores but are not intended for medical use or reviewed by the Food and Drug Administration (FDA). Our study compared oxygen saturation (SpO2) in patients from 1 medical use (MU) pulse oximeter to 8 NMU pulse oximeters. Measured arterial oxygenation (SaO2) was compared with SpO2 when available. In patients who were normoxic (SpO2 ≥90%), all oximeters exhibited similar readings. This finding suggests that NMU pulse oximeters may be able to rule out hypoxemia in clinical settings.Keywords
This publication has 6 references indexed in Scilit:
- The Accuracy of 6 Inexpensive Pulse Oximeters Not Cleared by the Food and Drug Administration: The Possible Global Public Health ImplicationsAnesthesia & Analgesia, 2016
- An assessment of the accuracy of pulse oximetersAnaesthesia, 2012
- Understanding and using sensitivity, specificity and predictive valuesIndian Journal Of Ophthalmology, 2008
- Statistics Notes: Diagnostic tests 2: predictive valuesBMJ, 1994
- History and recent developments in pulse oximetry.1993
- The validity and practicality of sun-reactive skin types I through VIArchives of Dermatology, 1988