Abstract
Purpose To assess the outcome in patients treated with proximal femoral locking compression plate (PF-LCP 4.5/5.0, Synthes©) for unstable inter- and subtrochanteric femoral fractures. Methods A retrospective analysis of 16 patients with proximal femur fractures (AO: 31A2: n = 5/32.3%; 31-A3: n = 10/62.5%; 32B1: n = 1/6.3%) treated with a PF-LCP at a Level 1 trauma centre between 2011 and 2015 was conducted. Results Sixteen patients were available for follow-up with a mean follow-up time of 14 months (range 4–29). Primary outcome included fracture healing, post-operative complications and post-operative ambulatory status. Male to female ratio was 1:1. Mean age was 61 ± 17 years. Union was achieved in a mean of 13.5 ± 3 weeks (range 12–20 weeks). Five patients (31.3%) had implant-associated complications like non-union, malrotation, late implant-associated infection, distal screw fractures and post-traumatic impingement of the hip. Consequently, four patients (25%) had to undergo revision surgery. There was no reported case of secondary varus collapse or cut-out. Conclusions Complications occurred in 31.3% (n = 5) in medium-term follow-up after PF-LCP in proximal unstable inter- and subtrochanteric femur fractures. These findings are supported by results of other groups. However, further studies to evaluate risk factors associated with failure of this implant are required.Keywords
This publication has 31 references indexed in Scilit:
- Radiographic union score for hip substantially improves agreement between surgeons and radiologistsBMC Musculoskeletal Disorders, 2013
- Failure of Proximal Femoral Locking Compression Plate: A Case SeriesJournal of Orthopaedic Trauma, 2011
- A Biomechanical Analysis of Locking Plate Fixation With Minimally Invasive Plate Osteosynthesis in a Subtrochanteric Fracture ModelJournal of Trauma: Injury, Infection & Critical Care, 2011
- A Biomechanical Comparison of Locked Plate Fixation With Percutaneous Insertion Capability Versus the Angled Blade Plate in a Subtrochanteric Fracture Gap ModelJournal of Orthopaedic Trauma, 2009
- Die instabile pertrochantäre Femurfraktur. Komplikationen, Fraktursinterung und Funktion nach extra- und intramedullärer Versorgung (PCCP™, DHS und PFN)Zeitschrift für Orthopädie und Unfallchirurgie, 2009
- The PFNA Proximal Femur Nail in Treatment of Unstable Proximal Femur Fractures-3 Cases of Postoperative Perforation of the Helical Blade Into the Hip JointJournal of Orthopaedic Trauma, 2008
- Functional comparison of the dynamic hip screw and the Gamma locking nail in trochanteric hip fractures: a matched-pair study of 268 patientsInternational Orthopaedics, 2007
- Evaluation of complications of three different types of proximal extra-articular femur fracturesInternational Orthopaedics, 2006
- Subtrochanteric fractures: Current management optionsDisability and Rehabilitation, 2005
- Dynamic Hip Screw With Trochanteric Stabilizing Plate in the Treatment of Unstable Proximal Femoral Fractures: A Comparative Study With the Gamma Nail and Compression Hip ScrewJournal of Orthopaedic Trauma, 1998