A phantom for dose-image quality optimization in chest radiography
Author:
- 1 October 2002
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 75 (898) , 837-842
Abstract
Optimization in chest radiography requires evaluation of patient dose and image quality. This study is aimed at proposing a simple geometrical phantom that realistically simulates the important anatomical regions of the thorax. For this purpose, the standard LucAl chest phantom is modified by adding an "anthropomorphic" insert and image quality test plate. Different test objects are arranged on the plate in three important anatomical areas; lung, cardiac, and subdiaphragmal regions. The aim is to simultaneously find two types of image quality index, objective and subjective, which can be used to compare different images in order to select the better image. Two objective indices are proposed, areal contrast index DeltaC(a) and scatter fraction P(s) and two subjectively estimated, a low contrast visualization index P(low) and a high contrast visualization index P(high). To demonstrate the potential of this phantom method it was applied to an X-ray unit to find the optical film density that ensures optimal visualization in different anatomical areas. It was found for the X-ray system under investigation that the automatic exposure control could be set to produce an optical density of about 1.8 in the lung field. The reported method is easily implemented in any clinical situation where optimization of chest radiography is needed.Keywords
This publication has 5 references indexed in Scilit:
- Results of a Nationwide Survey of Chest Radiography: Comparison with Results of a Previous StudyRadiology, 2000
- Quality control phantom for digital chest radiography.Radiology, 1997
- Chest radiography: estimated lung volume and projected area obscured by the heart, mediastinum, and diaphragm.Radiology, 1994
- Assessment and comparison of chest radiography techniques in the United States and SwedenBritish Journal of Radiology, 1990
- Beam quality independent attenuation phantom for estimating patient exposure from x-ray automatic exposure controlled chest examinationsMedical Physics, 1984