The relationship of pre-procedural Dmax based sizing to lesion level outcomes in Absorb BVS and Xience EES treated patients in the AIDA trial
Open Access
- 25 March 2019
- journal article
- research article
- Published by Springer Nature in The International Journal of Cardiovascular Imaging
- Vol. 35 (7) , 1189-1198
Abstract
Due to expansion limits of the Absorb bioresorbable scaffold a meticulous implantation with correct sizing is required. We sought to investigate the clinical outcomes based on the sizing of the device related to the maximal lumen diameter measured by quantitative coronary angiography in Absorb BVS and Xience EES treated lesions in the AIDA trial. Sizing of Absorb bioresorbable vascular scaffold (BVS) and Xience everolimus eluting stent (EES) was graded according to the definitions of device non-oversize and device oversize on pre-procedural angiography. Lesion-oriented outcomes (LOCE) (device thrombosis, TLR and TVMI) that occurred during 2 years follow-up were related to device non-oversized or oversized status. In the Absorb BVS group, LOCE occurred in 48 (7.4%) lesions in the oversized group and in 32 (8.2%) lesions in the non-oversized group (HR 0.91; 95% CI 0.58–1.42; p = 0.681), whereas TLR occurred in 34 (5.3%) lesions and in 23 lesions (5.9%), respectively (HR 0.89; 95% CI 0.52–1.51; p = 0.666). Definite scaffold thrombosis occurred in 11 (1.7%) device oversized treated lesions against 16 (4.1%) device non-oversized treated lesions (HR 0.41; 95% CI 0.19–0.89; p = 0.020). There were no differences in event rates between oversized and non-oversized groups in lesions treated with Xience EES. There was no significant difference in LOCE between oversized and non-oversized treated Absorb BVS and Xience EES treated lesions. Non-oversized Absorb BVS implantation was associated with a higher risk of scaffold thrombosis at complete 2 years follow-up. The majority of very late scaffold thrombosis occurred in properly sized devices.Keywords
Funding Information
- Abbott Vascular (n/a)
This publication has 17 references indexed in Scilit:
- 3-Year Clinical Outcomes With Everolimus-Eluting Bioresorbable Coronary ScaffoldsJACC, 2017
- Bioresorbable Scaffolds versus Metallic Stents in Routine PCINew England Journal of Medicine, 2017
- Dmax for sizing, PSP-1, PSP-2, PSP-3 or OCT guidance: interventionalist’s jargon or indispensable implantation techniques for short- and long-term outcomes of Absorb BRS?EuroIntervention, 2017
- Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trialThe Lancet, 2016
- The Impact of Post-Procedural Asymmetry, Expansion, and Eccentricity of Bioresorbable Everolimus-Eluting Scaffold and Metallic Everolimus-Eluting Stent on Clinical Outcomes in the ABSORB II TrialJACC: Cardiovascular Interventions, 2016
- Everolimus-Eluting Bioresorbable Scaffolds for Coronary Artery DiseaseNew England Journal of Medicine, 2015
- The ABSORB EXTEND study: preliminary report of the twelve-month clinical outcomes in the first 512 patients enrolledEuroIntervention, 2015
- Incidence and Imaging Outcomes of Acute Scaffold Disruption and Late Structural Discontinuity After Implantation of the Absorb Everolimus-Eluting Fully Bioresorbable Vascular ScaffoldJACC: Cardiovascular Interventions, 2014
- Dynamics of vessel wall changes following the implantation of the Absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study at 6, 12, 24 and 36 monthsEuroIntervention, 2014
- Five-Year Clinical and Functional Multislice Computed Tomography Angiographic Results After Coronary Implantation of the Fully Resorbable Polymeric Everolimus-Eluting Scaffold in Patients With De Novo Coronary Artery DiseaseJACC: Cardiovascular Interventions, 2013