Robust Catheter Tracking by Fusing Electromagnetic Tracking, Fiber Bragg Grating and Sparse Fluoroscopic Images
- 24 August 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 21 (20) , 23422-23434
Abstract
Endovascular catheterization is an intervention which offers a low risk alternative to open surgery in many patients. Today’s interventions rely heavily on fluoroscopic imaging to guide interventionalists. Fluoroscopy only produces 2D visualization of the catheter and also exposes both the patient and interventionalists to harmful radiation. Different approaches have been proposed to overcome the limitations of fluoroscopy. Fiber Bragg Grating (FBG)-based shape sensing is becoming popular to reconstruct the catheter shape. Multi-core fibers with parallel optical cores are interesting as they allow 3D shape reconstruction with a single fiber. A common issue with FBG-based shape sensing is its sensitivity to variations in twist. Even small amounts of twist can significantly impact the overall shape reconstruction accuracy. This work proposes a novel approach which combines electromagnetic tracking (EMT), FBG-based shape sensing, and sparse fluoroscopic images. The method provides real-time 3D visualization of the catheter without the need of continuous fluoroscopy. A unique feature of the proposed method is the selective use of imaging for dynamic twist-compensation of the FBG sensor. The proposed sensor-fusion method improved 3D reconstruction accuracy. Real-world in-vitro experiments promising results. For a catheter with an embedded fiber length of 170 mm, the proposed approach the 3D shape with a median root-mean-square (rms) error of 0.39 mm and an interquartile range of 0.10 mm in the 2D experiment in which the catheter was bent in a plane. A median rms error of 0.54 mm and an interquartile range of 0.07 mm were achieved in the 3D experiments.Keywords
All Related Versions
Funding Information
- ATLAS project
- European Union’s Horizon 2020 research and innovation program under the Marie Sklodowska-Curie grant agreement (813782)
- Internal KU Leuven C2 Project SF-REACTA
This publication has 29 references indexed in Scilit:
- 3D–2D registration for surgical guidance: effect of projection view angles on registration accuracyPhysics in Medicine & Biology, 2013
- Error Analysis of FBG-Based Shape Sensors for Medical Needle TrackingIEEE/ASME Transactions on Mechatronics, 2013
- The Use of Fiber Bragg Grating Sensors in Biomechanics and Rehabilitation Applications: The State-of-the-Art and Ongoing Research TopicsSensors, 2012
- Fiber grating sensors in medicine: Current and emerging applicationsSensors and Actuators A: Physical, 2011
- Impact of biplane versus single-plane imaging on radiation dose, contrast load and procedural time in coronary angioplastyBritish Journal of Radiology, 2010
- A flexible new technique for camera calibrationIEEE Transactions on Pattern Analysis and Machine Intelligence, 2000
- Multiscale vessel enhancement filtering1998
- Building Skeleton Models via 3-D Medial Surface Axis Thinning AlgorithmsCVGIP: Graphical Models and Image Processing, 1994
- Closed-form solution of absolute orientation using unit quaternionsJournal of the Optical Society of America A, 1987
- The Numerical Evaluation of B-SplinesIMA Journal of Applied Mathematics, 1972