In Vitro Performance Analysis of a Novel Pulsatile Diagonal Pump in a Simulated Pediatric Mechanical Circulatory Support System
- 14 November 2013
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 38 (1) , 64-72
Abstract
The objective of this study was to evaluate the pump performance of the third‐generation Medos diagonal pump, the Deltastream DP3, on hemodynamic profile and pulsatility in a simulated pediatric mechanical circulatory support (MCS) system. The experimental circuit consisted of a Medos Deltastream DP3 pump head and console (MEDOS Medizintechnik AG, Stolberg, Germany), a 14‐Fr Terumo TenderFlow Pediatric arterial cannula and a 20‐Fr Terumo TenderFlow Pediatric venous return cannula (Terumo Corporation, Tokyo, Japan), and 3 ft of tubing with an internal diameter of in. for both arterial and venous lines. Trials were conducted at flow rates ranging from 250 mL/min to 1000 mL/min (250‐mL/min increments) and rotational speeds ranging from 1000 to 4000 rpm (1000‐rpm increments) using human blood (hematocrit 40%). The postcannula pressure was maintained at 60 mm Hg by a Hoffman clamp. Real‐time pressure and flow data were recorded using a Labview‐based acquisition system. The pump provided adequate nonpulsatile and pulsatile flow, created more hemodynamic energy under pulsatile mode, and generated higher positive and negative pressures when the inlet and outlet of the pump head, respectively, were clamped. After the conversion from nonpulsatile to pulsatile mode, the flow rates and the rotational speeds increased. In conclusion, the novel Medos Deltastream DP3 diagonal pump is able to supply the required flow rate for pediatric MCS, generate adequate quality of pulsatility, and provide surplus hemodynamic energy output in a simulated pediatric MCS system.Keywords
This publication has 14 references indexed in Scilit:
- Hypovolemia in Extracorporeal Life Support Can Lead to Arterial Gaseous MicroemboliArtificial Organs, 2013
- Novel Pulsatile Diagonal Pump for Pediatric Extracorporeal Life Support SystemArtificial Organs, 2013
- Facts and Myths Surrounding Pediatric Mechanical Cardiovascular Circulatory Support Research: A Personal PerspectiveArtificial Organs, 2012
- Venovenous extracorporeal membrane oxygenation for acute lung failure in adultsThe Journal of Heart and Lung Transplantation, 2012
- Heart Transplantation for Congenital Heart DiseaseWorld Journal for Pediatric and Congenital Heart Surgery, 2011
- Mechanical Circulatory Support for End-Stage Heart Failure in Repaired and Palliated Congenital Heart DiseaseCurrent Cardiology Reviews, 2011
- New devices for pediatric mechanical circulatory supportCurrent Opinion in Cardiology, 2008
- Myths and Truths of Pulsatile and Nonpulsatile Perfusion During Acute and Chronic Cardiac SupportArtificial Organs, 2004
- Postcardiotomy assisted circulation with roller pump-early and late resultsActa chirurgica Iugoslavica, 2003
- Energy Equivalent PressureArchives of Surgery, 1966