Aiding the Detection of QRS Complex in ECG Signals by Detecting S Peaks Independently
Authors:
- 30 March 2018
- journal article
- research article
- Published by Springer Nature in Cardiovascular Engineering and Technology
- Vol. 9 (3) , 469-481
Abstract
In this paper, a novel algorithm for the accurate detection of QRS complex by combining the independent detection of R and S peaks, using fusion algorithm is proposed. R peak detection has been extensively studied and is being used to detect the QRS complex. Whereas, S peaks, which is also part of QRS complex can be independently detected to aid the detection of QRS complex. In this paper, we suggest a method to first estimate S peak from raw ECG signal and then use them to aid the detection of QRS complex. The amplitude of S peak in ECG signal is relatively weak than corresponding R peak, which is traditionally used for the detection of QRS complex, therefore, an appropriate digital filter is designed to enhance the S peaks. These enhanced S peaks are then detected by adaptive thresholding. The algorithm is validated on all the signals of MIT-BIH arrhythmia database and noise stress database taken from physionet.org. The algorithm performs reasonably well even for the signals highly corrupted by noise. The algorithm performance is confirmed by sensitivity and positive predictivity of 99.99% and the detection accuracy of 99.98% for QRS complex detection. The number of false positives and false negatives resulted while analysis has been drastically reduced to 80 and 42 against the 98 and 84 the best results reported so far.Keywords
Funding Information
- Department of Science and Technology, India (SR/WOS-A/ET-1049/2015(G))
This publication has 32 references indexed in Scilit:
- Novel Real-Time FPGA-Based R-Wave Detection Using Lifting WaveletCircuits, Systems, and Signal Processing, 2015
- QRS detection using S-Transform and Shannon energyComputer Methods and Programs in Biomedicine, 2014
- Straightforward and robust QRS detection algorithm for wearable cardiac monitorHealthcare Technology Letters, 2014
- Revisiting QRS Detection Methodologies for Portable, Wearable, Battery-Operated, and Wireless ECG SystemsPLOS ONE, 2014
- Denoising and R-Peak Detection of Electrocardiogram Signal Based on EMD and Improved Approximate EnvelopeCircuits, Systems, and Signal Processing, 2013
- R peak detection in electrocardiogram signal based on an optimal combination of wavelet transform, Hilbert transform, and adaptive thresholdingJournal of Medical Signals & Sensors, 2011
- A Wavelet-Based ECG Delineator: Evaluation on Standard DatabasesIEEE Transactions on Biomedical Engineering, 2004
- Using wavelet transforms for ECG characterization. An on-line digital signal processing systemIEEE Engineering in Medicine and Biology Magazine, 1997
- Detection of ECG characteristic points using wavelet transformsIEEE Transactions on Biomedical Engineering, 1995
- An approach to QRS complex detection using mathematical morphologyIEEE Transactions on Biomedical Engineering, 1993