Abstract
Isotope delta measurements are normalized against international reference standards. Although multi-point normalization is becoming a standard practice, the existing uncertainty evaluation practices are either undocumented or are incomplete. For multi-point normalization, we present errors-in-variables regression models for explicit accounting of the measurement uncertainty of the international standards along with the uncertainty that is attributed to their assigned values. This manuscript presents framework to account for the uncertainty that arises due to a small number of replicate measurements and discusses multi-laboratory data reduction while accounting for inevitable correlations between the laboratories due to the use of identical reference materials for calibration. Both frequentist and Bayesian methods of uncertainty analysis are discussed.Keywords
This publication has 20 references indexed in Scilit:
- Assessment of international reference materials for isotope-ratio analysis (IUPAC Technical Report)Pure and Applied Chemistry, 2014
- Normalization procedures and reference material selection in stable HCNOS isotope analyses: an overviewAnalytical and Bioanalytical Chemistry, 2012
- Calibration and uncertainty assessment for certified reference gas mixturesAnalytical and Bioanalytical Chemistry, 2010
- Laboratory effects models for interlaboratory comparisonsAccreditation and Quality Assurance, 2009
- Assessment of measurement uncertainty via observation equationsMetrologia, 2007
- Normalization of measured stable isotopic compositions to isotope reference scales – a reviewRapid Communications in Mass Spectrometry, 2007
- Atomic weights of the elements 2005 (IUPAC Technical Report)Pure and Applied Chemistry, 2006
- Combining information from interlaboratory evaluations using a random effects modelMetrologia, 2004
- Referencing strategies and techniques in stable isotope ratio analysisRapid Communications in Mass Spectrometry, 2001
- Guidelines for calibration in analytical chemistry. Part I. Fundamentals and single component calibration (IUPAC Recommendations 1998)Pure and Applied Chemistry, 1998