Pulse oximetry and high-dose vasopressors: a comparison between forehead reflectance and finger transmission sensors
- 7 August 2012
- journal article
- research article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 38 (10) , 1718-1722
Abstract
Pulse oximetry (SpO2) measured at finger site via transmission mode may fail in situations of hypoperfusion. Forehead sensors using reflectance technology might be useful in these circumstances. We hypothesized that reflectance SpO2 would be more accurate than finger SpO2 in patients with severe shock. A prospective observational study was conducted in an intensive care unit of a university hospital of patients in shock who were treated with high norepinephrine and/or epinephrine doses (≥0.1 μg kg−1 min−1). When blood gas determinations were requested, forehead SpO2 and finger SpO2 values were simultaneous recorded. Agreement between SpO2 measurements with arterial saturation (SaO2), obtained by blood analysis with a co-oximeter, was assessed using the Bland–Altman method. The number of outliers, defined by the formula SaO2 − SpO2 > ±3 %, indicated the proportion of measurements considered to be clinically unacceptable. Thirty-two patients were enrolled in the study. With the forehead sensor no reading failure occurred, and 140 paired data sets (forehead SpO2 vs. SaO2) were obtained. Bias and precision were +1.0 and +2.5 %, respectively, and the limits of agreement ranged from −4.0 to 6.0 %. The finger sensor failed to give a value in four cases, thus providing 136 paired data sets (finger SpO2 vs. SaO2) for analysis. Bias and precision were +1.4 and +4.8 %, respectively, and the limits of agreement ranged from −8.0 to 10.9 %. There were 21 (15 %) outliers for the forehead sensor and 43 (32 %) for the finger sensor (P < 0.001). Forehead SpO2 measurements were more accurate than finger SpO2 when compared with SaO2 in critically ill patients requiring high-dose vasopressor therapy and should therefore be the preferred method considered.Keywords
This publication has 22 references indexed in Scilit:
- Comparison of the Accuracy of Noninvasive Hemoglobin Monitoring by Spectrophotometry (SpHb) and HemoCue® with Automated Laboratory Hemoglobin MeasurementAnesthesiology, 2011
- The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort studyBMC Emergency Medicine, 2010
- A Comparison of Transmittance and Reflectance Pulse Oximetry During Vascular SurgeryAnesthesia & Analgesia, 2009
- Forehead Pulse OximetryAnesthesiology, 2006
- The desaturation response time of finger pulse oximeters during mild hypothermia*Anaesthesia, 2004
- Pulse oximetryIntensive Care Medicine, 2004
- Comparing and predicting between several methods of measurementBiostatistics, 2004
- Comparing and predicting between several methods of measurementBiostatistics, 2004
- Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unitCritical Care Medicine, 2000
- The accuracy of the biox 3700 pulse oximeter in patients receiving vasoactive therapyIntensive Care Medicine, 1991