A method to correct for stray light in telecentric optical-CT imaging of radiochromic dosimeters
Authors:
- Joseph NewtonJNJoseph Newton
- Mark OldhamMOMark Oldham
Open Access
- 30 June 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (14) , 4433-4451
Abstract
Radiochromic plastic and gel materials have recently emerged which can yield 3D dose information over clinical volumes in high resolution. These dosimeters can provide a much more comprehensive verification of complex radiation therapy treatments than can be achieved by conventional planar and point dosimeters. To achieve full clinical potential, these dosimeters require a fast and accurate read-out technology. Broad-beam optical-computed tomography (optical-CT) systems have shown promise, but can be sensitive to stray light artifacts originating in the imaging chain. In this work we present and evaluate a method to correct for stray light artifacts by deconvolving a measured, spatially invariant, point spread function (PSF). The correction was developed for the DLOS (Duke large field-of-view optical-CT scanner) in conjunction with radiochromic PRESAGE® dosimeters. The PSF was constructed from a series of acquisitions of projection images of various sized apertures placed in the optical imaging chain. Images were acquired with a range of exposure times, and for a range of aperture sizes (0.2-11 mm). The PSF is investigated under a variety of conditions, and found to be robust and spatially invariant, key factors enabling the viability of the deconvolution approach. The spatial invariance and robustness of the PSF are facilitated by telecentric imaging, which produces a collimated light beam and removes stray light originating upstream of the imaging lens. The telecentric capability of the DLOS therefore represents a significant advantage, both in keeping stray light levels to a minimum and enabling viability of an accurate PSF deconvolution method to correct for the residual. The performance of the correction method was evaluated on projection images containing known optical-density variations, and also on known 3D dose distributions. The method is shown to accurately account for stray light on small field dosimetry with corrections up to 3% in magnitude shown here although corrections of >10% have been observed in extreme cases. The dominant source of stray light was found to be within the imaging lens. Correcting for stray light extended the dynamic range of the system from ∼30 to ∼60 dB. The correction should be used when measurements need to be accurate within 3%.Keywords
This publication has 23 references indexed in Scilit:
- A method to correct for spectral artifacts in optical-CT dosimetryPhysics in Medicine & Biology, 2011
- Light scattering in optical CT scanning of Presage dosimetersJournal of Physics: Conference Series, 2010
- Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE®/ optical-CT 3D dosimetry systemPhysics in Medicine & Biology, 2010
- Three-dimensional dosimetry of small megavoltage radiation fields using radiochromic gels and optical CT scanningPhysics in Medicine & Biology, 2009
- Modelling optical scattering artefacts for varying pathlength in a gel dosimeter phantomPhysics in Medicine & Biology, 2008
- Focusing optics of a parallel beam CCD optical tomography apparatus for 3D radiation gel dosimetryPhysics in Medicine & Biology, 2006
- Optical-CT scanning of polymer gelsJournal of Physics: Conference Series, 2004
- Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: II. Optical properties of the BANG polymer gelPhysics in Medicine & Biology, 1996
- An iterative technique for the rectification of observed distributionsThe Astronomical Journal, 1974
- Bayesian-Based Iterative Method of Image Restoration*Journal of the Optical Society of America, 1972