A Novel Approach to Fabricate Early Keratoconus Phantom Models
Authors:
Open Access
- 1 April 2025
- journal article
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Translational Vision Science & Technology
- Vol. 14 (4) , 18
Abstract
To develop a method to fabricate early keratoconus phantom models and evaluate the feasibility of using corneal models for studying the dynamic response of early keratoconus under an air puff. A corneal mold was designed, and the silicone material was poured into the mold to produce corneal phantoms. Two types of early keratoconus phantoms with reduced mechanical properties in a specific area were prepared using a two-step molding process: the central keratoconus phantom and the paracentral keratoconus phantom. Corneal Visualization Scheimpflug Technology tonometry was performed on the normal corneal phantoms and early keratoconus phantoms, and the corresponding dynamic corneal response (DCR) parameters were recorded. A majority of DCR parameters of the normal corneal phantoms, including deflection amplitude at highest concavity (HCDA), peak distance (PD), radius of curvature (HR), first and second applanation times (A1T and A2T), first and second applanation velocities (A1V and A2V), and the stiffness parameter at the first applanation (SPA1), exhibited trends in response to changes in the simulated intraocular pressure (SIOP) that aligned with experimental results based on ex vivo animal eyes. Significant differences in HCDA, PD, HR, A1V, A2V, A1T, A2T, and integrated radius (IR) were observed between the early keratoconus phantoms and the normal corneal phantoms. The early keratoconus phantom models fabricated by the present novel approach are feasible for studying the dynamic response of early keratoconus under an air puff. This study demonstrated the potential of corneal phantom models for corneal biomechanical studies, which can deepen our understanding of the DCR parameters, and the results will provide valuable information for early diagnosis of keratoconus.Keywords
This publication has 47 references indexed in Scilit:
- Performance of Corvis ST Parameters Including Updated Stress-Strain Index in Differentiating Between Normal, Forme-Fruste, Subclinical, and Clinical Keratoconic EyesAmerican Journal of Ophthalmology, 2023
- Combinations of Scheimpflug tomography, ocular coherence tomography and air-puff tonometry improve the detection of keratoconusContact Lens and Anterior Eye, 2023
- Optimized Artificial Intelligence for Enhanced Ectasia Detection Using Scheimpflug-Based Corneal Tomography and Biomechanical DataAmerican Journal of Ophthalmology, 2022
- Multiscale mechanical model based on patient-specific geometry: Application to early keratoconus developmentJournal of the Mechanical Behavior of Biomedical Materials, 2022
- Keratoconus: An updated reviewContact Lens and Anterior Eye, 2022
- Ectasia After Corneal Refractive Surgery: A Systematic ReviewOphthalmology and Therapy, 2021
- Assessment of Corneal Biomechanical Properties with Inflation Test Using Optical Coherence TomographyAnnals of Biomedical Engineering, 2018
- Evaluation of the relationship of corneal biomechanical metrics with physical intraocular pressure and central corneal thickness in ex vivo rabbit eye globesExperimental Eye Research, 2015
- Cone Location–Dependent Outcomes After Combined Topography-Guided Photorefractive Keratectomy and Collagen Cross-linkingAmerican Journal of Ophthalmology, 2015
- Keratoconus: A reviewContact Lens and Anterior Eye, 2010