Improved volume selective 1H MR spectroscopic imaging of the prostate with gradient offset independent adiabaticity pulses at 3 tesla
Open Access
- 29 September 2014
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 74 (4) , 915-924
Abstract
Purpose: Volume selection in 1H MR spectroscopic imaging (MRSI) of the prostate is commonly performed with low‐bandwidth refocusing pulses. However, their large chemical shift displacement error (CSDE) causes lipid signal contamination in the spectral range of interest. Application of high‐bandwidth adiabatic pulses is limited by radiofrequency (RF) power deposition. In this study, we aimed to provide an MRSI sequence that overcomes these limitations.Methods: Measurements were performed at 3 T with an endorectal receive coil. A semi‐LASER sequence was equipped with low RF power demanding gradient‐modulated offset‐independent adiabaticity (GOIA) refocusing pulses with WURST(16,4) modulation, with a 10 kHz bandwidth.Results: Simulations and phantom studies verified that the GOIA pulses select slices with a flat top and sharp edges and minimal CSDE. The sequence timing was tuned to an optimal citrate signal shape at an echo time of 88 ms. Patient studies (n = 10) demonstrated that high quality spectra with reduced lipid artifacts can be obtained from the whole prostate. Compared with PRESS acquisition at 145 ms the signal‐to‐noise ratio (SNR) of citrate is increased up to 2.6 and choline up to 1.3.Conclusion: An MRSI sequence of the prostate is presented with minimized spectral lipid contamination and improved SNR, to facilitate routine clinical acquisition of metabolic data. Magn Reson Med 74:915–924, 2015.Keywords
Funding Information
- EU Marie Curie project FAST (MRTN-CT-2006-035801)
- The Dutch Cancer Society (KUN 2007/3971)
- Siemens Healthcare
This publication has 24 references indexed in Scilit:
- Mapping of prostate cancer by 1H MRSINMR in Biomedicine, 2013
- In vivo 1HMR spectroscopic imaging of aggressive prostate cancer: Can we detect lactate?Magnetic Resonance in Medicine, 2013
- Prostate Cancer: Multiparametric MR Imaging for Detection, Localization, and StagingRadiology, 2011
- Discriminating Cancer From Noncancer Tissue in the Prostate by 3-Dimensional Proton Magnetic Resonance Spectroscopic ImagingInvestigative Radiology, 2011
- Spectroscopic imaging with improved gradient modulated constant adiabaticity pulses on high-field clinical scannersJournal of Magnetic Resonance, 2010
- Short echo time 1H‐MRSI of the human brain at 3T with minimal chemical shift displacement errors using adiabatic refocusing pulsesMagnetic Resonance in Medicine, 2007
- Optimal timing for in vivo 1H‐MR spectroscopic imaging of the human prostate at 3TMagnetic Resonance in Medicine, 2005
- Fast acquisition-weighted three-dimensional proton MR spectroscopic imaging of the human prostateMagnetic Resonance in Medicine, 2004
- Adiabatic pulsesNMR in Biomedicine, 1997
- Adiabatic Pulses for Wideband Inversion and Broadband DecouplingJournal of Magnetic Resonance, Series A, 1995