MRI Meets MPI: A Bimodal MPI-MRI Tomograph

Authors:

Abstract
While magnetic particle imaging (MPI) constitutes a novel biomedical imaging technique for tracking superpara magnetic nanoparticles in vivo, unlike magnetic resonance imaging (MRI), it cannot provide anatomical background information. Until now these two modalities have been performed in separate scanners and image co-registration has been hampered by the need to reposition the sample in both systems as similarly as possible. This paper presents a bimodal MPI-MRI-tomograph that combines both modalities in a single system. MPI and MRI images can thus be acquired without moving the sample or replacing any parts in the setup. The images acquired with the presented setup show excellent agreement between the localization of the nano particles in MPI and the MRI background data. A combination of two highly complementary imaging modalities has been achieved.
Funding Information
  • German Federal Ministry of Education and Research (BMBF) (FKZ 1745X08)
  • IDEA project of the 7th framework programme of the European Union under Project 279288, and in part by the German Research Council (DFG) (BE 5293/1-1)

This publication has 21 references indexed in Scilit: