Amorphous silicon carbide probe mechanics for insertion in the cerebral cortex of rats, pigs, and macaques
Open Access
- 1 June 2025
- journal article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 22 (3) , 036029
Abstract
Objective. Intracortical microelectrode arrays (MEAs) are implantable devices used for neural recording and stimulation. However, their long-term performance is often compromised due, at least in part, to glial scar formation, initiated by microglial migration and astrocyte activation following implantation. To address this issue, ultra-thin MEAs (UMEAs) have been proposed as an alternative due to their reduced cross-sectional area (CSA) and enhanced flexibility that minimizes the mechanical mismatch between brain tissue and the electrode. These properties are expected to mitigate the persistent foreign-body response associated with micromotion. However, unaided implantation of UMEAs can be challenging, as their high flexibility may increase the likelihood of buckling and curtail the precise penetration into the brain. Approach. We investigated flexible amorphous silicon carbide (a-SiC) probe designs with varying CSAs and lengths to identify geometries that enable unsupported implantation into the cerebral cortices of rat, pig, and macaques. The critical buckling force of the a-SiC probes was experimentally determined as a function of geometry and formally described via finite element modeling, which predicted buckling behavior. Additionally, the penetration behavior of a-SiC probes was evaluated by measuring force–displacement responses during insertion into cortical tissues across species. Main results. Our findings demonstrated that the penetration force and cortical dimpling depth were not significantly influenced by the range of probe geometries tested. However, we observed that penetration force and dimpling depth were significantly lower in rat cortices than in larger species. Importantly, probe geometries with a higher ratio (⩾2) of critical buckling force to penetration force exhibited a 100% success rate for unaided insertion. Significance. This study provides a framework for designing and evaluating UMEA geometries to optimize unsupported implantation in both small and large animal brains.Keywords
Funding Information
- NIH (R01 NS104344)
This publication has 52 references indexed in Scilit:
- Design, fabrication, and implantation of invasive microelectrode arrays as in vivo brain machine interfaces: A comprehensive reviewJournal of Manufacturing Processes, 2024
- Planar amorphous silicon carbide microelectrode arrays for chronic recording in rat motor cortexBiomaterials, 2024
- A Critical Review of Microelectrode Arrays and Strategies for Improving Neural InterfacesAdvanced Healthcare Materials, 2019
- Transparent, Flexible, Penetrating Microelectrode Arrays with Capabilities of Single‐Unit ElectrophysiologyAdvanced Biosystems, 2019
- Effects of caspase-1 knockout on chronic neural recording quality and longevity: Insight into cellular and molecular mechanisms of the reactive tissue responseBiomaterials, 2014
- Neural probe design for reduced tissue encapsulation in CNSBiomaterials, 2007
- Plasma‐enhanced chemical vapor deposited silicon carbide as an implantable dielectric coatingJournal of Biomedical Materials Research Part A, 2003
- Biocompatibility of silicon‐based electrode arrays implanted in feline cortical tissueJournal of Biomedical Materials Research, 1993
- A method for pneumatically inserting an array of penetrating electrodes into cortical tissueAnnals of Biomedical Engineering, 1992
- Albumin and immunoglobulin in plasma and cerebrospinal fluid, and blood-cerebrospinal fluid barrier function in patients with dementia of alzheimer type and multi-infarct dementiaJournal of the Neurological Sciences, 1983