Dual-Energy CT: Lower Limits of Iodine Detection and Quantification
- 25 June 2019
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 292 (2) , 414-419
Abstract
The iodine limit of detection using dual-energy CT systems varies with scanner and phantom size, but all systems enable detection of iodine at or below 0.3 and 0.5 mg/mL in the small and large phan... BackgroundAssessments of the quantitative limitations among the six commercially available dual-energy (DE) CT acquisition schemes used by major CT manufacturers could aid researchers looking to us...Keywords
This publication has 22 references indexed in Scilit:
- Pseudoenhancement effects on iodine quantification from dual-energy spectral CT systems: A multi-vendor phantom study regarding renal lesion characterizationEuropean Journal of Radiology, 2018
- Objective image characterization of a spectral CT scanner with dual-layer detectorPhysics in Medicine & Biology, 2018
- Phase 1-2 Study of Dual-Energy Computed Tomography for Assessment of Pulmonary Function in Radiation Therapy PlanningInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- "How to" incorporate dual-energy imaging into a high volume abdominal imaging practice.Abdominal Radiology, 2017
- Dual-energy CT workflow: multi-institutional consensus on standardization of abdominopelvic MDCT protocolsAbdominal Radiology, 2016
- Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-FilterInvestigative Radiology, 2016
- Clinical significance of dual-energy CT-derived iodine quantification in the diagnosis of metastatic LN in colorectal cancerEuropean Journal of Surgical Oncology, 2015
- Iodine Quantification to Distinguish Clear Cell from Papillary Renal Cell Carcinoma at Dual-Energy Multidetector CT: A Multireader Diagnostic Performance StudyRadiology, 2014
- Interdependencies of acquisition, detection, and reconstruction techniques on the accuracy of iodine quantification in varying patient sizes employing dual-energy CTEuropean Radiology, 2014
- First performance evaluation of a dual-source CT (DSCT) systemEuropean Radiology, 2005