Digital CZT detector system for high flux energy-resolved X-ray imaging
- 1 October 2016
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-7
Abstract
Photon counting arrays with energy resolving capabilities are recently desired for the next-generation X-ray imaging systems. In this work, we present the performance of a 2 mm thick CZT pixel detector, with pixel pitches of 500 μm and 250 μm, coupled to a fast and low noise ASIC (PIXIE ASIC), characterized by only the preamplifier stage. A 16-channel digital readout electronics was used to continuously digitize and process each output channel from the PIXIE ASIC, performing multi-parameter analysis (event arrival time, pulse shape, pulse height) at low and high input counting rates (ICRs). The spectroscopic response of the system to monochromatic X-ray and gamma ray sources, at both low and high ICRs, is presented with particular attention to the mitigation of some typical spectral distortions (pile-up and charge sharing). The detector allows good energy resolution at moderate cooling (3% FWHM 59.5 keV, -1200 V, T = 5 C) by using fast shaped pulses (i.e. pulses with time widths of 300 ns). Charge sharing investigations were performed by using a fine time coincidence analysis (TCA) and a pulse shape analysis (PSA). The potentialities of the PSA for charge sharing detection even at high photon counting rates were also shown.Keywords
This publication has 36 references indexed in Scilit:
- X-ray response of CdZnTe detectors grown by the vertical Bridgman technique: Energy, temperature and high flux effectsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Electrical properties of Au/CdZnTe/Au detectors grown by the boron oxide encapsulated Vertical Bridgman techniqueNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Digital performance improvements of a CdTe pixel detector for high flux energy-resolved X-ray imagingNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015
- Interconnect and bonding techniques for pixelated X-ray and gamma-ray detectorsJournal of Instrumentation, 2015
- Real time digital pulse processing for X-ray and gamma ray semiconductor detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- Energy resolution and throughput of a new real time digital pulse processing system for x-ray and gamma ray semiconductor detectorsJournal of Instrumentation, 2013
- Electroless gold contact deposition on CdZnTe detectors by scanning pipette techniqueJournal of Instrumentation, 2012
- Performance enhancements of compound semiconductor radiation detectors using digital pulse processing techniquesNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- Study of Surface Treatment Effects on the Metal-CdZnTe InterfaceIEEE Transactions on Nuclear Science, 2009
- Characterization of Traveling Heater Method (THM) Grown ${\hbox{Cd}}_{0.9}{\hbox{Zn}}_{0.1}{\hbox{Te}}$ CrystalsIEEE Transactions on Nuclear Science, 2007