Estimation of organ and effective doses resulting from cone beam CT imaging for radiotherapy treatment planning
Authors:
- L J SawyerLSL J Sawyer
- S A WhittleSWS A Whittle
- E S MatthewsEME S Matthews
- H C StarrittHSH C Starritt
- T P JuppTJT P Jupp
- 1 July 2009
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 82 (979) , 577-584
Abstract
In this study, organ doses were measured for various kilovoltage cone beam CT exposures on the Varian Acuity simulator and an alternative method of dose estimation was also assessed. Organ doses were measured by distributing thermoluminescent dosimeters (TLDs) throughout an anthropomorphic phantom, and effective doses were calculated using International Commission on Radiological Protection (ICRP) 60 and ICRP 103 tissue-weighting factors. The ImPACT CT patient dosimetry calculator was also used to estimate doses for comparison with the TLD results. Effective doses of 15.3 mSv (19.4 mSv), 14.3 mSv (9.7 mSv) and 2.8 mSv (3.2 mSv) were calculated from the TLD measurements and ICRP 60 (ICRP 103) weighting factors for breast, pelvis and head acquisitions, respectively. When a 10 cm pencil ionisation chamber was used to measure the CT dose index, the ImPACT calculator was found to provide an adequate estimation of dose when compared with the TLD results. However, the doses for half-fan exposures were found to be overestimated, with the extent of overestimation depending on the radiosensitive organs irradiated. The organ and effective doses reported provide information for justification and optimisation of cone beam CT procedures, and are compared with doses delivered by other imaging devices. The ImPACT calculator may be used to estimate doses from cone beam CT procedures, if the potential for overestimation is acknowledged.Keywords
This publication has 9 references indexed in Scilit:
- Accurate patient dosimetry of kilovoltage cone‐beam CT in radiation therapyMedical Physics, 2008
- Doses to critical organs following radiotherapy and concomitant imaging of the larynx and breastBritish Journal of Radiology, 2007
- The management of imaging dose during image‐guided radiotherapy: Report of the AAPM Task Group 75Medical Physics, 2007
- Imaging doses from the Elekta Synergy X-ray cone beam CT systemBritish Journal of Radiology, 2007
- Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancerPhysics in Medicine & Biology, 2007
- Patient dose and image quality from mega‐voltage cone beam computed tomography imagingMedical Physics, 2007
- Organ doses from prostate radiotherapy and associated concomitant exposuresBritish Journal of Radiology, 2006
- Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapyMedical Physics, 2006
- Enlarged longitudinal dose profiles in cone‐beam CT and the need for modified dosimetryMedical Physics, 2005