An MLC calibration method using a detector array
Authors:
- Thomas A. SimonTSThomas A. Simon
- Darren KahlerDKDarren Kahler
- William E. SimonWSWilliam E. Simon
- Christopher FoxCFChristopher Fox
- Jonathan LiJLJonathan Li
- Jatinder PaltaJPJatinder Palta
- Chihray LiuCLChihray Liu
- 9 September 2009
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 36 (10) , 4495-4503
Abstract
The authors have developed a quantitative calibration method for a multileaf collimator (MLC) which measures individual leaf positions relative to the MLC backup jaw on an Elekta Synergy linear accelerator. The method utilizes a commercially available two-axis detector array (Profiler 2; Sun Nuclear Corporation, Melbourne, FL). To calibrate the MLC bank, its backup jaw is positioned at the central axis and the opposing jaw is retracted to create a half-beam configuration. The position of the backup jaws field edge is then measured with the array to obtain what is termed the radiation defined reference line. The positions of the individual leaf ends relative to this reference line are then inferred by the detector response in the leaf end penumbra. Iteratively adjusting and remeasuring the leaf end positions to within specifications completes the calibration. Using the backup jaw as a reference for the leaf end positions is based on three assumptions: (1) The leading edge of an MLC leaf bank is parallel to its backup jaw's leading edge, (2) the backup jaw position is reproducible, and (3) the measured radiation field edge created by each leaf end is representative of that leaf's position. Data from an electronic portal imaging device (EPID) were used in a similar analysis to check the results obtained with the array. The relative leaf end positions measured with the array differed from those measured with the EPID by an average of 0.11+/-0.09 mm per leaf. The maximum leaf positional change measured with the Profiler 2 over a 3 month period was 0.51 mm. A leaf positional accuracy of +/-0.4 mm is easily attainable through the iterative calibration process. The method requires an average of 40 min to measure both leaf banks. This work demonstrates that the Profiler 2 is an effective tool for efficient and quantitative MLC quality assurance and calibration.Keywords
This publication has 9 references indexed in Scilit:
- MLC quality assurance using EPID: A fitting technique with subpixel precisionMedical Physics, 2008
- Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancerPhysics in Medicine & Biology, 2007
- A dosimetric calibration method for a double‐focused multileaf collimatorMedical Physics, 2007
- Evaluation of MLC leaf positioning using a scanning liquid ionization chamber EPIDPhysics in Medicine & Biology, 2006
- Transmission, Penumbra and Leaf Positional Accuracy in Commissioning and Quality Assurance Program of a Multileaf Collimator for Step-and-Shoot IMRT TreatmentsTumori Journal, 2006
- Evaluation of two methods of predicting MLC leaf positions using EPID measurementsMedical Physics, 2006
- A Fourier analysis of the dose grid resolution required for accurate IMRT fluence map optimizationMedical Physics, 2005
- Dosimetric considerations for a multileaf collimator systemMedical Physics, 1996
- The design and performance characteristics of a multileaf collimatorPhysics in Medicine & Biology, 1994