Analitical Study of Chirp Pulse Microwave Breast Radar (CP-MBR)
- 1 December 2007
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21654727,p. 1-4
Abstract
Feasibility of a new microwave radar imaging system for early stage breast tumor detection using chirp pulse microwaves was investigated numerically. In the FDTD-based simulation, the basic idea of discriminating tumors from surrounding inhomogeneous breast tissues was verified and the effects of the bandwidth of CP-MBR on the resolution were examined by varying the frequency band from 2-3 GHz to 1-4 GHz.Keywords
This publication has 8 references indexed in Scilit:
- An overview of ultra-wideband microwave imaging via space-time beamforming for early-stage breast-cancer detectionIEEE Antennas and Propagation Magazine, 2005
- Image generation in chirp pulse microwave computed tomography (CP-MCT) by numerical computation: Computation of a human head modelElectronics and Communications in Japan (Part III: Fundamental Electronic Science), 2005
- Visualization of the breast tumor by the integrated use of CP-MCT and chirp pulse microwave breast radar2005
- Microwave imaging via space-time beamforming for early detection of breast cancerIEEE Transactions on Antennas and Propagation, 2003
- Experimental validation of a linear model for data reduction in chirp-pulse microwave CTIEEE Transactions on Medical Imaging, 2002
- Image restoration in chirp-pulse microwave CT (CP-MCT)IEEE Transactions on Biomedical Engineering, 2000
- Two-dimensional FDTD analysis of a pulsed microwave confocal system for breast cancer detection: fixed-focus and antenna-array sensorsIEEE Transactions on Biomedical Engineering, 1998
- Tomographic measurement of temperature change in phantoms of the human body by chirp radar-type microwave computed tomographyMedical & Biological Engineering & Computing, 1993