Biometry and intraocular lens power calculation results with a new optical biometry device: Comparison with the gold standard
Authors:
- 1 April 2014
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 40 (4) , 593-600
Abstract
To evaluate the agreement in axial length (AL), keratometry (K), anterior chamber depth (ACD) measurements; intraocular lens (IOL) power calculations; and predictability using a new partial coherence interferometry (PCI) optical biometer (AL-Scan) and a reference (gold standard) PCI optical biometer (IOLMaster 500). Service d'Ophtalmologie, Hopital Bicêtre, APHP Université, Paris, France. Evaluation of a diagnostic device. One eye of consecutive patients scheduled for cataract surgery was measured. Biometry was performed with the new biometer and the reference biometer. Comparisons were performed for AL, average K at 2.4 mm, ACD, IOL power calculations with the Haigis and SRK/T formulas, and postoperative predictability of the devices. A P value less than 0.05 was statistically significant. The study enrolled 50 patients (mean age 72.6 years±4.2 SEM). There was a good correlation between biometers for AL, K, and ACD measurements (r=0.999, r=0.933, and r=0.701, respectively) and between IOL power calculation with the Haigis formula (r=0.972) and the SRK/T formula (r=0.981). The mean absolute error (MAE) in IOL power prediction was 0.42±0.08 diopter (D) with the new biometer and 0.44±0.08 D with the reference biometer. The MAE was 0.20 D with the Haigis formula and 0.19 with the SRK/T formula (P=.36). The new PCI biometer provided valid measurements compared with the current gold standard, indicating that the new device can be used for IOL power calculations for routine cataract surgery. No author has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 18 references indexed in Scilit:
- Comparison of 2 laser instruments for measuring axial lengthJournal of Cataract & Refractive Surgery, 2010
- Intraocular lens power calculation using the IOLMaster and various formulas in eyes with long axial lengthJournal of Cataract & Refractive Surgery, 2008
- Comparison of the Zeiss IOLMaster and applanation A‐scan ultrasound: biometry for intraocular lens calculationClinical & Experimental Ophthalmology, 2003
- Immersion A-scan compared with partial coherence interferometryJournal of Cataract & Refractive Surgery, 2002
- Improved prediction of intraocular lens power using partial coherence interferometryJournal of Cataract & Refractive Surgery, 2001
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Partial coherence interferometry: a novel approach to biometry in cataract surgeryAmerican Journal of Ophthalmology, 1998
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992
- Comparison of the SRK/T formula and other theoretical and regression formulasJournal of Cataract & Refractive Surgery, 1990
- A three-part system for refining intraocular lens power calculationsJournal of Cataract & Refractive Surgery, 1988