An in vitro visual study of fugitive aerosols released during aerosol therapy to an invasively ventilated simulated patient
Open Access
- 1 January 2021
- journal article
- research article
- Published by Taylor & Francis in Drug Delivery
- Vol. 28 (1) , 1496-1500
Abstract
COVID-19 can cause serious respiratory complications resulting in the need for invasive ventilatory support and concurrent aerosol therapy. Aerosol therapy is considered a high risk procedure for the transmission of patient derived infectious aerosol droplets. Critical-care workers are considered to be at a high risk of inhaling such infectious droplets. The objective of this work was to use noninvasive optical methods to visualize the potential release of aerosol droplets during aerosol therapy in a model of an invasively ventilated adult patient. The noninvasive Schlieren imaging technique was used to visualize the movement of air and aerosol. Three different aerosol delivery devices: (i) a pressurized metered dose inhaler (pMDI), (ii) a compressed air driven jet nebulizer (JN), and (iii) a vibrating mesh nebulizer (VMN), were used to deliver an aerosolized therapeutic at two different positions: (i) on the inspiratory limb at the wye and (ii) on the patient side of the wye, between the wye and endotracheal tube, to a simulated intubated adult patient. Irrespective of position, there was a significant release of air and aerosol from the ventilator circuit during aerosol delivery with the pMDI and the compressed air driven JN. There was no such release when aerosol therapy was delivered with a closed-circuit VMN. Selection of aerosol delivery device is a major determining factor in the release of infectious patient derived bioaerosol from an invasively mechanically ventilated patient receiving aerosol therapy.Keywords
This publication has 22 references indexed in Scilit:
- Clinical course and outcomes of critically ill patients with SARS-CoV-2 pneumonia in Wuhan, China: a single-centered, retrospective, observational studyThe Lancet Respiratory Medicine, 2020
- Baseline Characteristics and Outcomes of 1591 Patients Infected With SARS-CoV-2 Admitted to ICUs of the Lombardy Region, ItalyJAMA, 2020
- The Confusing World of Dry Powder Inhalers: It Is All About Inspiratory Pressures, Not Inspiratory Flow RatesJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2020
- Investigation of Fugitive Aerosols Released into the Environment during High-Flow TherapyPharmaceutics, 2019
- Aerosol therapy during mechanical ventilation: an international surveyIntensive Care Medicine, 2013
- Some aspects of the airborne transmission of infectionJournal of The Royal Society Interface, 2009
- Albuterol Delivery in a Model of Mechanical Ventilation. Comparison of Metered-Dose Inhaler and Nebulizer EfficiencyAmerican Journal of Respiratory and Critical Care Medicine, 1995
- Evaluation of a Reservoir Device for Metered-Dose Bronchodilator Delivery to Intubated AdultsChest, 1992
- Metered Dose Inhaler Aerosol Characteristics Are Affected by the Endotracheal Tube Actuator/Adapter UsedAnesthesiology, 1990
- AERODYNAMICS OF THE HUMAN MICROENVIRONMENTThe Lancet, 1969