Noninvasive Average Flow Estimation for an Implantable Rotary Blood Pump: A New Algorithm Incorporating the Role of Blood Viscosity
- 21 December 2006
- journal article
- Published by Wiley in Artificial Organs
- Vol. 31 (1) , 45-52
Abstract
Abstract: The effect of blood hematocrit (HCT) on a noninvasive flow estimation algorithm was examined in a centrifugal implantable rotary blood pump (iRBP) used for ventricular assistance. An average flow estimator, based on three parameters, input electrical power, pump speed, and HCT, was developed. Data were collected in a mock loop under steady flow conditions for a variety of pump operating points and for various HCT levels. Analysis was performed using three‐dimensional polynomial surfaces to fit the collected data for each different HCT level. The polynomial coefficients of the surfaces were then analyzed as a function of HCT. Linear correlations between estimated and measured pump flow over a flow range from 1.0 to 7.5 L/min resulted in a slope of 1.024 L/min (R2 = 0.9805). Early patient data tested against the estimator have shown promising consistency, suggesting that consideration of HCT can improve the accuracy of existing flow estimation algorithms.Keywords
This publication has 10 references indexed in Scilit:
- First Clinical Implant of the VentrAssist Left Ventricular Assist System as Destination Therapy for End-Stage Heart FailureThe Journal of Heart and Lung Transplantation, 2005
- Measurement for implantable rotary blood pumpsPhysiological Measurement, 2005
- In Vivo Test of Pressure Head and Flow Rate Estimation in a Continuous‐Flow Artificial HeartArtificial Organs, 2003
- Non-invasive flow estimation in an implantable rotary blood pump: a study considering non-pulsatile and pulsatile flowsPhysiological Measurement, 2003
- Flow Rate and Pressure Head Estimation in a Centrifugal Blood PumpArtificial Organs, 2002
- Sensorless Estimation of Pressure Head and Flow of a Continuous Flow Artificial Heart Based on Input Power and Rotational SpeedASAIO Journal, 2002
- Sensorless Flow and Head Estimation in the VentrAssist Rotary Blood PumpArtificial Organs, 2000
- Physical Model‐Based Indirect Measurements of Blood Pressure and Flow Using a Centrifugal PumpArtificial Organs, 2000
- Noninvasive Pump Flow Estimation of a Centrifugal Blood PumpArtificial Organs, 1997
- An Evaluation of Thermodilution Cardiac Output Measurement Using the Swan-Ganz CatheterAnaesthesia and Intensive Care, 1981