Analysis of the Role of Lead Resistivity in Specific Absorption Rate for Deep Brain Stimulator Leads at 3T MRI
- 22 March 2010
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 29 (4) , 1029-1038
Abstract
Magnetic resonance imaging (MRI) on patients with implanted deep brain stimulators (DBSs) can be hazardous because of the antenna-effect of leads exposed to the incident radio-frequency field. This study evaluated electromagnetic field and specific absorption rate (SAR) changes as a function of lead resistivity on an anatomically precise head model in a 3T system. The anatomical accuracy of our head model allowed for detailed modeling of the path of DBS leads between epidermis and the outer table. Our electromagnetic finite difference time domain (FDTD) analysis showed significant changes of 1 g and 10 g averaged SAR for the range of lead resistivity modeled, including highly conductive leads up to highly resistive leads. Antenna performance and whole-head SAR were sensitive to the presence of the DBS leads only within 10%, while changes of over one order of magnitude were observed for the peak 10 g averaged SAR, suggesting that local SAR values should be considered in DBS guidelines. With ¿lead = ¿copper , and the MRI coil driven to produce a whole-head SAR without leads of 3.2 W/kg, the 1 g averaged SAR was 1080 W/kg and the 10 g averaged SAR 120 W/kg at the tip of the DBS lead. Conversely, in the control case without leads, the 1 g and 10 g averaged SAR were 0.5 W/kg and 0.6 W/kg, respectively, in the same location. The SAR at the tip of lead was similar with electrically homogeneous and electrically heterogeneous models. Our results show that computational models can support the development of novel lead technology, properly balancing the requirements of SAR deposition at the tip of the lead and power dissipation of the system battery.Keywords
This publication has 45 references indexed in Scilit:
- Electromagnetic and modeling analyses of an implanted device at 3 and 7 TeslaJournal of Magnetic Resonance Imaging, 2007
- The Characterization of Conductive Textile Materials Intended for Radio Frequency ApplicationsIEEE Antennas and Propagation Magazine, 2007
- EEG/(f)MRI measurements at 7 Tesla using a new EEG cap (“InkCap”)NeuroImage, 2006
- MRI‐induced heating of selected thin wire metallic implants – laboratory and computational studies – findings and new questions raisedMinimally Invasive Therapy & Allied Technologies, 2006
- Resistive Tapered Stripline (RTS) in Electroencephalogram Recordings During MRIIEEE Transactions on Microwave Theory and Techniques, 2004
- An algorithm for computations of the power deposition in human tissueIEEE Antennas and Propagation Magazine, 1999
- Applied Bioelectricity1998
- Analysis of Miniature Electric Field Probes with Resistive Transmission LinesIEEE Transactions on Microwave Theory and Techniques, 1981
- Design and calibration of the NBS isotropic electric-field monitor (EFM-5), 0.2 to 1000 MHz1981
- Numerical solution of initial boundary value problems involving maxwell's equations in isotropic mediaIEEE Transactions on Antennas and Propagation, 1966