Appropriate Face Models for Evaluating Drug Delivery in the Laboratory: The Current Situation and Prospects for Future Advances
- 1 March 2008
- journal article
- research article
- Published by SAGE Publications in Journal of Aerosol Medicine and Pulmonary Drug Delivery
- Vol. 21 (1) , 97-112
Abstract
The laboratory evaluation of inhalers with facemasks for patient interface is so complex that testing without a facemask is generally undertaken, a practice that has been advocated in one standard. However, the facemask itself can profoundly influence medication delivery. A systematic review of the literature was undertaken to establish the development history of face models for the evaluation of facemasks used with inhalers and accessories. Initial attempts to simulate the facemask–face boundary employed a circular, firm rubber flange plate upon which the facemask was located. However, such models did not represent dead volume accurately, which is particularly important when assessing infant use. Subsequent developments included the creation of more realistic facial features, enabling the aerosol leaving the inhaler to be quantified at the facemask. In one instance (SAINT model), an anatomically correct nasopharyngeal cavity has been combined with a model face, enabling assessment of medication delivery to be extended to the lower respiratory tract. However, it is necessary either to apply sealants or to compress the facemask beyond normal to eliminate leakage with the rigid facial structure that is incomplete above the bridge of the nose. An oral-breathing infant full-face model (ADAM) intended to be used to quantify emitted mass at the patient interface incorporates flexible facial features to overcome this limitation. There is a need to extend the flexible face approach to other models that may be developed in the future for testing facemasks, whether or not they incorporate anatomically correct realizations of the upper respiratory tract.Keywords
This publication has 41 references indexed in Scilit:
- Variation in Pediatric Aerosol Delivery: Importance of FacemaskJournal of Aerosol Medicine, 2005
- Canadian Standards Association Standard CAN/CSA/Z264.1‐02:2002: A New Voluntary Standard for Spacers and Holding Chambers Used with Pressurized Metered‐Dose InhalersCanadian Respiratory Journal, 2004
- Aerosol Therapy and the Fighting Toddler: Is Administration During Sleep an Alternative?Journal of Aerosol Medicine, 2003
- Considerations for the Development and Practice of Cascade Impaction Testing, Including a Mass Balance Failure Investigation TreeJournal of Aerosol Medicine, 2003
- Ethical aspects of using radiolabelling in aerosol researchArchives of Disease in Childhood, 2003
- Inhalation therapy for infantsAdvanced Drug Delivery Reviews, 2003
- Aerosol Therapy With Valved Holding Chambers in Young Children: Importance of the Facemask SealPediatrics, 2001
- European Respiratory Society Guidelines on the use of nebulizersEuropean Respiratory Journal, 2001
- Aerosol delivery from spacers in wheezy infants: a daily life studyEuropean Respiratory Journal, 2000
- Impact of Oropharyngeal Deposition on Inhaled DoseJournal of Aerosol Medicine, 1998