Polymer gel dosimetry for the TG-43 dosimetric characterization of a new125I interstitial brachytherapy seed
Authors:
- P PapagiannisPPP Papagiannis
- E PantelisEPE Pantelis
- E GeorgiouEGE Georgiou
- P KaraiskosPKP Karaiskos
- A AngelopoulosAAA Angelopoulos
- L SakelliouLSL Sakelliou
- S StiliarisSSS Stiliaris
- D BaltasDBD Baltas
- I SeimenisISI Seimenis
- 3 April 2006
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (8) , 2101-2111
Abstract
In this work, a polymer gel-magnetic resonance (MR) imaging method is employed for the dosimetric characterization of a new 125I low dose rate seed (IsoSeed model I25.S17). Two vials filled with PABIG gel were prepared in-house and one new seed as well as one commercially available 125I seed of similar dose rate and well-known dosimetric parameters (IsoSeed model I25.S06) were positioned in each vial. Both seeds in each vial were MR scanned simultaneously on days 11 and 26 after implantation. The data obtained from the known seed in each vial are used to calibrate the gel dose response which, for the prolonged irradiation duration necessitated by the investigated dose rates, depends on the overall irradiation time. Data for this study are presented according to the AAPM TG-43 dosimetric formalism. Polymer gel results concerning the new seed are compared to corresponding, published dosimetric results obtained, for the purpose of the new seed clinical implementation, by our group using the established methods of Monte Carlo (MC) simulation and thermo-luminescence dosimetry (TLD). Polymer gel dosimetry yields an average dose rate constant value of lambda = (0.921 +/- 0.031) cGy h(-1) U(-1) relative to (MC)lambda = (0.929 +/- 0.014) cGy h(-1) U(-1), (TLD)lambda = (0.951 +/- 0.044) cGy h(-1) U(-1) and the average value of Lambda = (0.940 +/- 0.051) cGy h(-1) U(-1) proposed for the clinical implementation of the new seed. Results for radial dose function, g(L)(r), and anisotropy function, F(r, theta), also agree with corresponding MC calculations within experimental uncertainties which are smaller for the polymer gel method compared to TLD. It is concluded that the proposed polymer gel-magnetic resonance imaging methodology could be used at least as a supplement to the established techniques for the dosimetric characterization of new low energy and low dose rate interstitial brachytherapy seeds.Keywords
This publication has 16 references indexed in Scilit:
- An evaluation of the TSE MR sequence for time efficient data acquisition in polymer gel dosimetry of applications involving high doses and steep dose gradientsMedical Physics, 2005
- Monte Carlo and thermoluminescence dosimetry of the new model I25.S17 interstitial brachytherapy seedMedical Physics, 2005
- Polymer gel dosimetry close to an125I interstitial brachytherapy seedPhysics in Medicine & Biology, 2005
- Polymer gel water equivalence and relative energy response with emphasis on low photon energy dosimetry in brachytherapyPhysics in Medicine & Biology, 2004
- Polymer gel dosimetry using a three‐dimensional MRI acquisition techniqueMedical Physics, 2002
- On the accuracy of monomer/polymer gel dosimetry in the proximity of a high-dose-rate192Ir sourcePhysics in Medicine & Biology, 2001
- Monte Carlo dosimetry of the selectSeed interstitial brachytherapy seedMedical Physics, 2001
- Artefacts in multi-echoT2imaging for high-precision gel dosimetry: II. Analysis ofB1-field inhomogeneityPhysics in Medicine & Biology, 2000
- Monte Carlo‐aided dosimetry of the symmetra model I25.S06 interstitial brachytherapy seedMedical Physics, 2000
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995