Characteristics and outcomes of gastrointestinal bleeding in patients with continuous‐flow left ventricular assist devices: A systematic review
- 16 May 2020
- journal article
- review article
- Published by Wiley in Artificial Organs
- Vol. 44 (11) , 1150-1161
Abstract
Gastrointestinal bleeding (GIB) is a common adverse event after continuous‐flow left ventricular assist device (CF‐LVAD) implantation. We sought to evaluate patterns of GIB development and related outcomes in CF‐LVAD recipients. An electronic search was performed to identify all articles related to GIB in the setting of CF‐LVAD implantation. A total of 34 studies involving 1087 patients were pooled for analysis. Mean patient age was 60 years (95% CI 57‐64) and 24% (95% CI 21‐28%) were female. The mean time from CF‐LVAD implantation to the first GIB was 54 days (95% CI 24‐84) with 40% (95% CI 34‐45%) of patients having multiple episodes of GIB. Anemia was present in 75% (95% CI 41‐93%) and the most common etiology of bleeding was arteriovenous malformations (36% [95% CI 24‐50%]). The mean duration of follow‐up was 14.6 months (95% CI 6.9‐22.3) during which the all‐cause mortality rate was 21% (95% CI 12‐36%) and the mortality rate from GIB was 4% (95% CI 2‐9%). Thromboembolic events occurred in 32% (95% CI 22‐44%) of patients with an ischemic stroke rate of 16% (95% CI 3‐51%) and a pump thrombosis rate of 8% (95%CI 3‐22%). Heart transplantation was performed in 31% (95% CI 18‐47%) of patients, after which 0% (95% CI 0‐10%) experienced recurrent GIB. GIB is a major source of morbidity among CF‐LVAD recipients. While death due to GIB is rare, cessation of anticoagulation during treatment increases the risk of subsequent thrombotic events. Heart transplant in these patients appears to reliably resolve the risk of future GIB.Keywords
This publication has 46 references indexed in Scilit:
- Clinical Differences Between Continuous Flow Ventricular Assist Devices: A Comparison Between HeartMate II and HeartWare HVADJournal of Cardiac Surgery, 2013
- Pulsatility and the Risk of Nonsurgical Bleeding in Patients Supported With the Continuous-Flow Left Ventricular Assist Device HeartMate IICirculation: Heart Failure, 2013
- Acquired von Willebrand disease in Jarvik 2000 recipients: A single center experienceInternational Journal of Cardiology, 2012
- Gastrointestinal bleeding with the HeartMate II left ventricular assist deviceThe Journal of Heart and Lung Transplantation, 2012
- Acquired von Willebrand Syndrome After Continuous-Flow Mechanical Device Support Contributes to a High Prevalence of Bleeding During Long-Term Support and at the Time of TransplantationJACC, 2010
- Management Options to Treat Gastrointestinal Bleeding in Patients Supported on Rotary Left Ventricular Assist Devices: A Single‐Center ExperienceArtificial Organs, 2010
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceNew England Journal of Medicine, 2009
- Gastrointestinal bleeding rates in recipients of nonpulsatile and pulsatile left ventricular assist devicesThe Journal of Thoracic and Cardiovascular Surgery, 2009
- Gastrointestinal bleeding from arteriovenous malformations in patients supported by the Jarvik 2000 axial-flow left ventricular assist deviceThe Journal of Heart and Lung Transplantation, 2005
- Quantifying heterogeneity in a meta‐analysisStatistics in Medicine, 2002