Novel Real-Time Fluoroscopic Assessment Method of THV Expansion Following Balloon-Expandable TAVR
- 1 February 2025
- journal article
- Published by Wolters Kluwer Health in Circulation: Cardiovascular Interventions
- Vol. 18 (2) , e014617
Abstract
BACKGROUND: Transcatheter heart valve (THV) underexpansion after transcatheter aortic valve replacement may be associated with worse outcomes. THV expansion can be assessed fluoroscopically using a pigtail for calibration; however, the accuracy of this technique specific to transcatheter aortic valve replacement is unknown. We assessed the accuracy and reproducibility of a novel fluoroscopic method to assess THV expansion using the THV commissural post for calibration. METHODS: Patients who underwent transcatheter aortic valve replacement with a SAPIEN 3 (S3) THV had documented 3 cusp and cusp overlap views, and post-implant computed tomography was identified. THV expansion was fluoroscopically assessed in the 3 cusp and cusp overlap views using the S3 commissural post height and the pigtail as a reference for calibration. The correlation between the 2 methods and computed tomography was evaluated. RESULTS: Forty patients were included. On the bench, the commissural post height measured 3.3, 3.5, 4.0, and 4.5 mm for the 20 mm S3, 23 mm S3, 26 mm S3, and 29 mm S3, respectively. The Pearson correlation coefficient (r) with computed tomography for measuring the inflow, mid-portion, and outflow THV diameter was 0.98, 0.97, and 0.98 for the commissural post height method and 0.82, 0.81, and 0.78 for the pigtail method, respectively. Unlike the pigtail method, the correlation between the commissural post height method and computed tomography remained strong across all THV sizes and in both the 3 cusp and cusp overlap views. CONCLUSIONS: The commissural post height method is a novel real-time fluoroscopic tool that has the potential to assess THV expansion and guide further optimization after balloon-expandable transcatheter aortic valve replacement.Keywords
This publication has 10 references indexed in Scilit:
- Bioprosthetic Valve Fracture 3 Years Post–Valve-in-Valve TAVRJACC: Cardiovascular Interventions, 2023
- Deformation of Transcatheter Aortic Valve Prostheses: Implications for Hypoattenuating Leaflet Thickening and Clinical OutcomesCirculation, 2022
- Late Balloon Valvuloplasty for Transcatheter Heart Valve DysfunctionJACC, 2022
- Incomplete expansion of transcatheter aortic valves is associated with propensity for valve thrombosisInteractive CardioVascular and Thoracic Surgery, 2019
- Computed Tomography Imaging in the Context of Transcatheter Aortic Valve Implantation (TAVI)/Transcatheter Aortic Valve Replacement (TAVR)JACC: Cardiovascular Imaging, 2019
- Subclinical leaflet thickening and stent frame geometry in self-expanding transcatheter heart valvesEuroIntervention, 2017
- Transcatheter Valve Underexpansion Limits Leaflet Durability: Implications for Valve-in-Valve ProceduresAnnals of Biomedical Engineering, 2016
- Total ellipse of the heart valve: the impact of eccentric stent distortion on the regional dynamic deformation of pericardial tissue leaflets of a transcatheter aortic valve replacementJournal of The Royal Society Interface, 2015
- Evaluation of catheters and metallic catheter markers as calibration standard for measurement of coronary dimensionCatheterization and Cardiovascular Diagnosis, 1990
- Use of a metal ring‐marked catheter for geometric calibration in quantitative coronary angiographyCatheterization and Cardiovascular Diagnosis, 1988